5.4.3 Focus Considerations Accurate measurement requires proper focus of the image. The part image is at best focus when the magnification is highest. It is often preferable to first focus the image at highest magnification, then decrease the magnification to the desired level. 5.4.4 Parts Fixturing The part must be fixtured securely to prevent part movement during measurement. Options are available from Starrett Kinemetric for off-the-shelf, semi-custom or custom fixturing. Please contact Starrett Kinemetric sales with your requirements. Aligning the part’s X or Y axis to the stage will improve dimensional measurements. If the part is off-axis from the stage, X-Y-Z measurements will not correlate as well with true part dimensions. Orientation error can also be removed by creating a reference frame based on the part before creating measurements. Please see the M3 Software Manual for details on the skew function.