MVR Manual

2 Environmental, Regulatory, and Safety Information

This section provides the regulatory, environmental, and safety information for the MVR systems.

2.1 Environmental Conditions

Operating Conditions

MVR systems can be safely operated under the following environmental conditions:

Environmental Condition

Operational Requirement

Ambient Temperature

20°C ± 3°C (68°F ± 5°F)

Humidity

< 90% RH

If the system is to be operated under environmental conditions that are substantially different from those in the previous table, the system should be recalibrated under the expected conditions. Users should also consider the material characteristics of the parts under inspection, in particular coefficients of thermal expansion. Numerical compensation might be required when measuring parts under conditions different from those controlling the stated dimensional specifications for these parts.

Factory Calibration Conditions

MVR systems are factory calibrated under the standard laboratory environmental conditions shown below:

Specification

Calibration Requirement

Ambient Temperature

20°C ± 0.5°C (68°F ± 5°F)

Humidity

40-60% RH

Temperature rate of change

0.5°C (1°F) per hour

2.2 Regulatory Compliance

The MVR200 and MVR300 Video Metrology Systems have been inspected, tested, and evaluated by independent test laboratories and are declared to comply with Council Directives 2006/42/EC - Machinery and 2002/95/EC RoHS (by exemption) when installed and operated in accordance with this manual. Accordingly, these systems are entitled to bear the CE Mark. The Product Safety, EMC Testing and Evaluation reports can be provided upon request.