■ The facility did not ensure that reverse osmosis systems were meeting the requirements of AAMI 5.2.2, Table 1 when tested with the typical feed water. » Location 1 - The RO 2 product conductivity range is 1-20 mS. On several dates it was documented out of range (22- 24). Additionally, the rejection range is >90%, however on several dates there is no documentation of this parameter. » Location 2 - The RO prefilter differential delta pressure range is <10. It was noted as “10” on nine occasions in June, nine occasions in July, and 10 occasions in August with no documentation at either location of trending analysis or interventions regarding the documented out of range measurements.

Compliance tips for:

The facility must be able to demonstrate that water and equipment used for dialysis meets quality standards and requirements from in the Association for the Advancement of Medical Instrumentation publication “Dialysate for hemodialysis,” ANSI/AAMI RD52:2004. ■ Audit water and dialysate logs for full and accurate documentation of requirements. ■ Conduct semi-annual training with relevant staff. ■ Hang laminated reminder signage in the appropriate areas to prompt staff on acceptable ranges and documentation requirements.

Nerd Newbies (understand the requirement)

Nerd Apprentices (audit for excellence)

Nerd Trailblazers (prepare the path for others)

RD7-J Frequency of the citation: 33%

Overview of the requirement: The treatment environment is designed, equipped, and maintained to be safe, functional, and comfortable for patients, staff, and visitors. Comment on deficiencies:  Compliance is assessed primarily through direct observation. Surveyor comments focused on the layout of patient care areas and deferred maintenance of equipment that compromised safety. Examples of ACHC Surveyor findings: ■ Patient rooms in this LTC facility are semi-private with two to three patients located in each room. During the survey, there were two patients dialyzing at the same time in each room. The waste lines/tubing were run from the machine into the restroom, then were taped to the toilet seat for waste disposal. Facility staff entering and exiting the rooms multiple times were carefully watching their footing. The tubing is not taped or secured to the floor creating a clear tripping hazard.

achc.org | (855) 937-2242 | 15

Made with FlippingBook. PDF to flipbook with ease