Clinical Benefits & The Löwenstein Difference

S A N T È • LÖW E N S T E I N S L E E P T H E R A PY D E V I C E S

Appendix 2 Pgs. 13-16

TwoAPAPSettings You have a choice between two different settings for APAP management.

next generation sourcing

Standard-APAP: AlgorithmDetails • An epoch = 2 minutes • Goal: Acclimation phase to a new pressure, no provocation of events or arousals • In general: pressure change at the end of the epoch • In event of obstructive hypopnea (oH) & obstructive apnea (oA), up to two times in the epoch • eSO events: oA, oH Snoring > eight breaths in the epoch • eMO events: RERA, Snoring > four breaths in the epoch

next generation sourcing

Made with FlippingBook Online newsletter