L2 Plus Creating a Test

3.3.1.5 Tensile to Home Step The Tensile to Home position requires that you establish a Home position prior to starting your test. This can be done simply by pressing the Set Home target or by establishing Home automatically when you start the test by configuring the Pre Test Set Home option to Yes. This step will move the crosshead to the established Home position.

These are the Exceptions for this step. An exception is an event. An exception stops the test. Exceptions are optional, but are recommended. Load Limit - If the crosshead moves and the measured load equals this limit, the crosshead will stop. Min Break - This is the minimum load that must first be achieved before the system begins to look for a Break % drop in load.

Enter the Speed. This is the velocity the crosshead will move at to the Home position.

When Collect Data = Yes, the individual data points for this test step are acquired and saved for exporting. When Collect Data = No, no data points are collected or saved. The step is just a “goto” without any data collected.

L2 Plus Test Builder - Tensile Step Types

L2 Plus Test Builder - Go to Home Moves the crosshead to the established Home position