L2 Plus Analyzing Your Test

5.7.2.4 Minimum Algorithm Where The WHERE function is used to specify where you want the Minimum result to display. Display the point on any of these views: • Results View • Graph Views using a Marker • Data Views (a column within the table)

5.7.2.5 Minimum Algorithm Tolerance Your Minimum result (coefficient) can make use of a tolerance to dis- tinguish the result as either a “pass” or “fail” relative to your tolerance limits. When you use a tolerance, you specify two tolerance limits. If your result equals or falls within the range created by these two tolerances, the result is displayed in black text and is deemed a “pass”. If the result falls outside the tolerance range, the result displays in red text and is deemed a “fail”. Your L2 Plus Series software also features a Tolerance view (TOL) that displays a tendency graph for the result. This graph displays whether the actual result is at the high or low end of the tolerance range you have established.

The coefficient Lpeak has a tolerance range between 4.00 and 5.00N. Results equal to or within this range display black indicating an “in-tolerance” result. Results outside this range display red indicating an “out-of-tolerance” result.

Minimum Peak with Tolerance Out-of-tolerance results display in red text

34

Made with FlippingBook - professional solution for displaying marketing and sales documents online