ENGLISH

4.2.6.2 BATCH PROMPT The Batch Prompt requires the operator performing the test to enter the name of the batch or the sample being tested. When the Batch Prompt is used, the operator is required to enter the name of the sample being tested prior to the test being able to be performed. The Batch Prompt default is OFF (the combo box is not highlighted). 4.2.7 SET DATUM OPTION The Set Datum option lets you configure your test setup to automatically zero the Load and/or Distance when the test is started- when START TEST is pressed. 4.2.7.1 ZERO LOAD If you select the combo box for Zero Load, the load measurement in the display is automatically set to a 0 value at the start of the test- when START TEST is pressed. 4.2.7.2 ZERO DISTANCE If you also select the combo box for the Zero Distance, your distance measurement in the display is automatically set to a 0 value at the start of the test- when START TEST is pressed.

4.2.6.1 OPERATOR PROMPT The Operator Prompt requires the operator performing the test to enter their name when the START TEST touch target is touched. Use this prompt so that the name of the operator performing the test is recorded in the test data. The Operator Prompt provides an audit trail for each test run by operator name. The Operator Prompt default is OFF (the combo box is not highlighted).

Shown: L1 Pre Test Operator Prompt setup

Shown: L1 Pre Test Batch Number Prompt setup

PKG08807 - UML1 25