ENGLISH
4.3.4.4 TARGET 1 Target 1 represents the first target. It may be either Load or Distance based on the Limit Type setup. If the Type=Tension, the crosshead will move upward until it achieves the Target 1 value. It will then move next to Target 2. If the Type=Compression, the crosshead will move downward until it achieves the Target 1 value. It will then move next to Target 2. 4.3.4.5 DWELL 1 You may configure a "pause duration", called a dwell for each Target 1 and Target 2. Dwell 1 is the amount of time, in seconds, that the crosshead will hold at the Target 1 limit. Dwell time may be set to up to 3600 seconds (60 minutes). When the crosshead moves to the target, the crosshead will hold at this position for the dwell duration. At the conclusion of the dwell duration, the crosshead will move next to Target 2. The default for Dwell is null. When null, there is no pause. The crosshead will move to Target 1, then reverse direction to Target 2. NOTE The L1 system can save up to 100,000 data points for a Test Run. When using any type of hold or cycle duration, you must use an appropriate Sampling Rate (Hz) so that the entire data collection for the Test Run doesn't exceed 100,000. 4.3.4.6 SPEED 1 The Speed 1 is the crosshead velocity to Target 1. 4.3.4.7 TARGET 2 Target 2 represents the second target. It may be either Load or Distance based on the Limit Type setup. If the Type=Tension, the crosshead will move upward until it achieves the Target 1 value. It will then move next to Target 2. If the Type=Compression, the crosshead will move downward until it achieves the Target 1 value. It will then move next to Target 2. 4.3.4.8 DWELL 2 You may configure a "pause duration", called a dwell for each Target 1 and Target 2. Dwell 2 is the amount of time, in seconds, that the crosshead will hold at the Target 2 limit. Dwell time may be set to up to 3600 seconds (60 minutes). When the crosshead moves to the target, the crosshead will hold at this position for the dwell duration. At the conclusion of the dwell duration, the crosshead will move next to Target 1. The default for Dwell is null. When null, there is no pause. The crosshead will move to Target 2, then reverse direction to Target 1. NOTE The L1 system can save up to 100,000 data points for a Test Run. When using any type of hold or cycle duration, you must use an appropriate Sampling Rate (Hz) so that the entire data collection for the Test Run doesn't exceed 100,000.
4.3.4.3 METHOD TYPE There are two cycle methods: Count or Duration. If the Method Type=Count, the test will cycle continuously for the number of counts, up to 99,999 cycles. If the Method Type=Duration, the test will cycle continuously for a time duration, up to 99,999 seconds.
Shown: L1 CYCLE COUNT Test Method setup
Shown: L1 CYCLE DURATION Test Method setup
32 PKG08807 - UML1
Made with FlippingBook - professional solution for displaying marketing and sales documents online