4.0 TEST STEPS All test setups have four fundamental steps: • Pre Test • Step Moves • Data (Coefficients) • Post Test

The Load Units may be either N (newtons), LBF (pound-force), KGF (kilogram-force), GF (gram-force) and OZF (ounce-force). Some higher capacity load cell sensors will not allow GF or OZF units to be selected due to the available display resolutions. The Distance Units may be either in (inches) or mm (millimeters). If Set Home = YES, the location where the crosshead is positioned when your test is started is automatically assigned as the Home position. The factory setting for Sampling (Hz) is 100 or 100 samples per second. During a test, data is sampled and collected 100 times per second. You have the option to increase or decrease the sampling rate from 5 to 1,000. The Last Batch Step option is usefull when you are performing the same test setup on multiple samples or parts ( a "batch" of parts). The ASK step may be used (at the beginning of the test setup) to ask questions that only need to be answered once, such as "Operator Name" or "Batch ID". These ASK steps will be skipped on the second and subsequent runs of the test if the Last Batch Step is set to the first step after them.

The Pre Test step lets you define activities and functions that occur prior to the actual testing, hence the term, Pre Test. Pre Test options include:

• Lock Test • Load Units • Distance Units • Set Home • Sampling (Hz) • Last Batch Step • Load Cell • Height Mode

Lock Test may be either NO (disabled) or YES (enabled). If enabled, the test setup cannot be revised, edited, or deleted unless performed by an authorized user.

Pre Test menu

Data (Coefficients) menu

Pre Test options

Post Test menu

Crosshead Move types