7.0 STEPS AND DATA GATHERING Steps have an option called Data Gathering. When Data Gathering = YES, data is collected by the system for the move. This data is used for graphing and available for export as raw data. When the Data Gathering = NO, no data is collected. The move changes to a "go to" step. A Go To move collects no data. A Go To step simply causes the crosshead to move without collecting data to your designated target. You can create a preconditioning step in your test setup using a Go To step.

When Collect Data = YES, your system will collect the raw data associated with the move step.

A "Go To" step symbol appears differently than a move step where data is being collected. The symbol is "unfilled" to indicate a "Go To".

Expand or collapse (close) some sections within the test setup menus. The "+" expands the selection. The "-" closes or collapses the selection.

When Collect Data = NO, your system will not collect raw data associated with the move. This move is now a "Go To" and causes the crosshead to move without collecting data during the move procedure.

Sections within an expanded step may be closed or expanded, including the Data Gathering section. Data gathering is always on by default in the Tension, Compression and Hold steps. Data Gathering does not exist in the Cycle, Datum or Ask steps. Data is not saved for these steps.

PKG08819-UML2SOFT 19