ENGLISH
9.2 DISTANCE HOLD You may hold the crosshead at a Distance target to determine the effects the hold time has on the sample's load characteristic. The Distance Hold is performed together with an associated tension or compress to distance step. The maximum allowable hold duration is 15 minutes (900 seconds). • Select the test mode- either Tension or Compression. • Enter the Distance Limit (target).
• Enter the Test Speed. • Select the Hold symbol.
• Select the Distance Hold symbol. • Enter the Hold Duration value.
In our example, we are NOT GATHERING DATA during the Hold period.
In our example, we setup a Compressive Distance Limit. The Compressive Hold Move is used to hold the load for 15 minutes.