L2 User Manual

ENGLISH

17.1.5 DATA SUMMARY VIEW The Data Summary View is a batch report for an entire test setup. The Data Summary View looks like a spreadsheet. Each row of information corresponds to a RUN within the test (batch). The Title Row shows the various titles, which include the coefficients and their units of measure. If Tolerances are used, a "tendency graph" is shown depicting the actual result on a line graph relative to the Limit 1 and Limit 2 range. Results that fall outside the tolerance range are displayed in RED text. The Shared function is listed in the footer. Selecting this function will create and send a .csv file of the Raw Data for a result.

The Title Row shows the Run No., coefficient results with units of measure and tolerance limits (if used). The highlighted (blue) row corresponds to the highlighted Run number in the Runs List.

When an "out-of-tolerance" condition occured, the RUN displays RED indicating a "failed" result.

The Tolerance Limits are displayed so you know the Limit 1 and Limit 2 values.

Shows active Data Summary View

When tolerance is used, a Tendency graph is used to depict how the measured result compares to the limit range you had established for Limit 1 and Limit 2. Our example shows that we exceeded the Limit 2 (Upper) tolerance.

Exports Raw Data for this RUN in a .csv file format