Calibration Certificate Certificate No. 746710 225001 Mountz Cal ID:

www.mountztorque.com

# AC-1346

Calibration Performed By: MOUNTZ ALABAMA 19051 UNDERWOOD RD. FOLEY, AL 36535 USA (408) 850-9978

Issued To:

19051 UNDERWOOD RD. FOLEY, AL 36535 USA MOUNTZ ALABAMA

Sales Order Number: SO268939

Equipment Tested:

Cal. Date:

6 MONTHS 08-Nov-2025 08-May-2025

Description:

MOUNTZ EZIII 50I TORQUE ANALYZER ELECTRONIC

Cal. Interval:

Arrival Condition: Cal. Due Date:

Serial Number: Model Number:

2208110002

MEETS MFG SPEC

EZIII 50I

F Service Performed: CAL TO MFG SPEC / Temp. / RH%: 69 48

Range:

5 - 50

Unit of Measure: lbf.in

Calibration Result:

0.13% PASS

Accuracy:

1% OF READING FROM 10 - 20% FS 0.5% OF READING FROM 20 - 100% FS

Uncertainty: Procedure:

3WI-7.1-003

Calibration Notes:

Measurement Data:

Seq.

Description

Nominal

Tolerance -

Tolerance +

As Found As Left

1 2 3 4 5 6 7 8 9

CW 10% Reading 1 CW 10% Reading 2 CW 10% Reading 3 CW 50% Reading 1 CW 50% Reading 2 CW 50% Reading 3 CW 100% Reading 1 CW 100% Reading 2 CW 100% Reading 3 CCW 10% Reading 1 CCW 10% Reading 2 CCW 10% Reading 3 CCW 50% Reading 1 CCW 50% Reading 2 CCW 50% Reading 3 CCW 100% Reading 1 CCW 100% Reading 2 CCW 100% Reading 3

4.994 4.994 4.994 24.97 24.97 24.97 49.94 49.94 49.94 4.994 4.994 4.994 24.97 24.97 24.97 49.94 49.94 49.94

4.944 4.944 4.944 24.85 24.85 24.85 49.69 49.69 49.69 4.944 4.944 4.944 24.85 24.85 24.85 49.69 49.69 49.69

5.044 5.044 5.044 25.10 25.10 25.10 50.19 50.19 50.19 5.044 5.044 5.044 25.10 25.10 25.10 50.19 50.19 50.19

4.992 4.989 4.996 24.97 24.96 24.97 49.99 49.98 49.99 5.008 5.006 5.011 25.01 25.00 24.99 50.03 50.02 50.01

4.992 4.989 4.996 24.97 24.96 24.97 49.99 49.98 49.99 5.008 5.006 5.011 25.01 25.00 24.99 50.03 50.02 50.01

10 11 12 13 14 15 16 17 18

Standards Used:

Description

I.D.

Range

Units Accuracy

Cal. Due

Traceability #

WEIGHT SET 1 CLASS F PRIMARY STANDARD

10040 0.016 - 50

lb in

0.01% 0.05%

27-Sep-2025 09-Jan-2026

45435-081061

MOUNTZ W4.0 PRIMARY STANDARD RADIUS WHEEL 38299 4

597561

Work Performed By: C. SALERNO TECHNICIAN

Reviewed By: CALIBRATION QC T. HOLLEY

Calibrations are performed using standards traceable through NIST or other National Metrology Institute to the International System of Units (SI). Calibration system complies with ANSI/NCSL Z540-1 and ISO17025. This information applies only to the instrument identified above and may not be reproduced, except in full, without prior written consent. Reported uncertainties are expressed as expanded values at approximately 95% confidence level using coverage factor K=2. There is no implied warranty that instrument will maintain its specified tolerance during the calibration interval. Decision Rule 1 is employed (measurement uncertainty is not taken into account) when making conformity statements unless otherwise specified .

Page 1 of 1

Calibration Certificate Certificate No. 740301 169841 Mountz Cal ID:

www.mountztorque.com

# AC-1346

Calibration Performed By: MOUNTZ ALABAMA 19051 UNDERWOOD RD. FOLEY, AL 36535 USA (408) 850-9978

Issued To:

19051 UNDERWOOD RD. FOLEY, AL 36535 USA MOUNTZ ALABAMA

Sales Order Number: SO266691

Equipment Tested:

Cal. Date:

12 MONTHS 13-Mar-2026 13-Mar-2025

Description:

MOUNTZ BMX2500F TORQUE TRANSDUCER FIXED

Cal. Interval:

Arrival Condition: Cal. Due Date:

Serial Number: Model Number:

BMX1907047 BMX2500F 250 - 2,500

MEETS MFG SPEC

F Service Performed: CAL TO MFG SPEC / Temp. / RH%: 70 42

Range:

Unit of Measure: lbf.ft

Calibration Result:

0.15% PASS

Accuracy:

1% OF READING FROM 10 - 20% FS 0.5% OF READING FROM 20 - 100% FS

Uncertainty: Procedure:

3WI-7.1-004

Calibration Notes:

Calibrated using PTT s/n: PTT0607118 version 2.5.2.5. rev 1.26. Gain checked. A/D count reading: +155

Measurement Data:

Seq.

Description

Nominal

Tolerance -

Tolerance +

As Found As Left

1 2 3 4 5 6 7 8 9

CW 10% Reading 1 CW 10% Reading 2 CW 10% Reading 3 CW 50% Reading 1 CW 50% Reading 2 CW 50% Reading 3 CW 100% Reading 1 CW 100% Reading 2 CW 100% Reading 3 CCW 10% Reading 1 CCW 10% Reading 2 CCW 10% Reading 3 CCW 50% Reading 1 CCW 50% Reading 2 CCW 50% Reading 3 CCW 100% Reading 1 CCW 100% Reading 2 CCW 100% Reading 3

250.00 250.00 250.00 1250.00 1250.00 1250.00 2500.00 2500.00 2500.00 250.00 250.00 250.00 1250.00 1250.00 1250.00 2500.00 2500.00 2500.00

247.50 247.50 247.50 1243.75 1243.75 1243.75 2487.50 2487.50 2487.50 247.50 247.50 247.50 1243.75 1243.75 1243.75 2487.50 2487.50 2487.50

252.50 252.50 252.50 1256.25 1256.25 1256.25 2512.50 2512.50 2512.50 252.50 252.50 252.50 1256.25 1256.25 1256.25 2512.50 2512.50 2512.50

250.87 250.61 250.78 1247.38 1247.22 1247.10 2498.71 2498.66 2498.54 250.86 250.77 250.68 1249.63 1249.34 1249.57 2497.33 2497.11 2497.49

250.87 250.61 250.78 1247.38 1247.22 1247.10 2498.71 2498.66 2498.54 250.86 250.77 250.68 1249.63 1249.34 1249.57 2497.33 2497.11 2497.49

10 11 12 13 14 15 16 17 18

Standards Used:

Description

I.D.

Range

Units Accuracy

Cal. Due

Traceability #

1111HL-10K TORQUE TRANSDUCER LOAD CELL

10001 100 - 10000

lbf.ft

0.02%

25-Mar-2025 22-Jul-2026

S000003429

HP 34401-A MULTIMETER DIGITAL

73677 0 - 99

mV/V 0.0015%

19012428

Work Performed By: T. HOLLEY TECHNICIAN

Reviewed By: CALIBRATION QC A. RHODES

Calibrations are performed using standards traceable through NIST or other National Metrology Institute to the International System of Units (SI). Calibration system complies with ANSI/NCSL Z540-1 and ISO17025. This information applies only to the instrument identified above and may not be reproduced, except in full, without prior written consent. Reported uncertainties are expressed as expanded values at approximately 95% confidence level using coverage factor K=2. There is no implied warranty that instrument will maintain its specified tolerance during the calibration interval. Decision Rule 1 is employed (measurement uncertainty is not taken into account) when making conformity statements unless otherwise specified .

Page 1 of 1

Calibration Certificate Certificate No. 746673 131973 Mountz Cal ID:

www.mountztorque.com

# AC-1346

Calibration Performed By: MOUNTZ ALABAMA 19051 UNDERWOOD RD. FOLEY, AL 36535 USA (408) 850-9978

Issued To:

19051 UNDERWOOD RD. FOLEY, AL 36535 USA MOUNTZ ALABAMA

Sales Order Number: SO268939

Equipment Tested:

Cal. Date:

6 MONTHS 08-Nov-2025 08-May-2025

Description:

MOUNTZ BMX20Z TORQUE TRANSDUCER FIXED

Cal. Interval:

Arrival Condition: Cal. Due Date:

Serial Number: Model Number:

BMX1211033

OUT OF MFG SPEC

BMX20Z

F Service Performed: CAL TO MFG SPEC / Temp. / RH%: 70 40

Range:

2 - 20

Unit of Measure: ozf.in

Calibration Result:

ADJ.

Accuracy:

1% OF READING FROM 10 - 20% FS 0.5% OF READING FROM 20 - 100% FS

Uncertainty: Procedure:

0.13%

3WI-7.1-003

Calibration Notes:

Calibrated using PTT s/n: 2206004 version 2.5.2.5. rev 1.26. Gain checked. A/D -102

Measurement Data:

Seq.

Description

Nominal

Tolerance -

Tolerance +

As Found As Left

1 2 3 4 5 6 7 8 9

CW 10% Reading 1 CW 10% Reading 2 CW 10% Reading 3 CW 50% Reading 1 CW 50% Reading 2 CW 50% Reading 3 CW 100% Reading 1 CW 100% Reading 2 CW 100% Reading 3 CCW 10% Reading 1 CCW 10% Reading 2 CCW 10% Reading 3 CCW 50% Reading 1 CCW 50% Reading 2 CCW 50% Reading 3 CCW 100% Reading 1 CCW 100% Reading 2 CCW 100% Reading 3

1.9976 1.9976 1.9976 9.9882 9.9882 9.9882 19.9763 19.9763 19.9763 1.9976 1.9976 1.9976 9.9882 9.9882 9.9882 19.9763 19.9763 19.9763

1.9777 1.9777 1.9777 9.9382 9.9382 9.9382 19.8764 19.8764 19.8764 1.9777 1.9777 1.9777 9.9382 9.9382 9.9382 19.8764 19.8764 19.8764

2.0176 2.0176 2.0176 10.0381 10.0381 10.0381 20.0762 20.0762 20.0762 2.0176 2.0176 2.0176 10.0381 10.0381 10.0381 20.0762 20.0762 20.0762

A A A A A A A A A A A A A

2.2025 2.2076 2.2054 10.2122 10.2151 10.2166 20.4274 20.4432 20.4497 1.6166 1.6173 1.6188 9.9322 9.9416 9.9395 19.9812 19.9950 19.9928

1.9812 1.9814 1.9835 9.9452 9.9473 9.9489 20.0118 20.0216 20.0230 1.9809 1.9816 1.9823 9.9639 9.9667 9.9681 20.0072 20.0016 20.0009

10 11 12 13 14 15 16 17 18

Standards Used:

Description

I.D.

Range

Units Accuracy

Cal. Due

Traceability #

WEIGHT SET 1 CLASS F PRIMARY STANDARD

10040 0.016 - 50

lb in

0.01%

27-Sep-2025 15-Feb-2026

45435-081061

MOUNTZ W1.0 PRIMARY STANDARD RADIUS WHEEL 250335 1

0.1%

692580

Work Performed By: R. COUCH TECHNICIAN

Reviewed By: CALIBRATION QC T. HOLLEY

Calibrations are performed using standards traceable through NIST or other National Metrology Institute to the International System of Units (SI). Calibration system complies with ANSI/NCSL Z540-1 and ISO17025. This information applies only to the instrument identified above and may not be reproduced, except in full, without prior written consent. Reported uncertainties are expressed as expanded values at approximately 95% confidence level using coverage factor K=2. There is no implied warranty that instrument will maintain its specified tolerance during the calibration interval. Decision Rule 1 is employed (measurement uncertainty is not taken into account) when making conformity statements unless otherwise specified .

Page 1 of 1

Calibration Certificate Certificate No. 743309 214726 Mountz Cal ID:

www.mountztorque.com

# AC-1346

Calibration Performed By: MOUNTZ ALABAMA 19051 UNDERWOOD RD. FOLEY, AL 36535 USA (408) 850-9978

Issued To:

19051 UNDERWOOD RD. FOLEY, AL 36535 USA MOUNTZ ALABAMA

Sales Order Number: SO267742

Equipment Tested:

Cal. Date:

07-Apr-2025

Description:

MOUNTZ EZIII 10I TORQUE ANALYZER ELECTRONICI

Cal. Interval:

6 MONTHS 07-Oct-2025

Arrival Condition: Cal. Due Date:

Serial Number: Model Number:

2111110007

MEETS MFG SPEC

EZIII 10I

F Service Performed: CAL TO MFG SPEC / Temp. / RH%: 70 64

Range:

1 - 10

Unit of Measure: lbf.in

Calibration Result:

0.13% PASS

Accuracy:

1% OF READING FROM 10 - 20% FS 0.5% OF READING FROM 20 - 100% FS

Uncertainty: Procedure:

3WI-7.1-003

Calibration Notes:

Measurement Data:

Seq.

Description

Nominal

Tolerance -

Tolerance +

As Found As Left

1 2 3 4 5 6 7 8 9

CW 10% Reading 1 CW 10% Reading 2 CW 10% Reading 3 CW 50% Reading 1 CW 50% Reading 2 CW 50% Reading 3 CW 100% Reading 1 CW 100% Reading 2 CW 100% Reading 3 CCW 10% Reading 1 CCW 10% Reading 2 CCW 10% Reading 3 CCW 50% Reading 1 CCW 50% Reading 2 CCW 50% Reading 3 CCW 100% Reading 1 CCW 100% Reading 2 CCW 100% Reading 3

0.999 0.999 0.999 4.994 4.994 4.994 9.988 9.988 9.988 0.999 0.999 0.999 4.994 4.994 4.994 9.988 9.988 9.988

0.989 0.989 0.989 4.969 4.969 4.969 9.938 9.938 9.938 0.989 0.989 0.989 4.969 4.969 4.969 9.938 9.938 9.938

1.009 1.009 1.009 5.019 5.019 5.019

0.999 1.001 0.999 4.992 4.993 4.991 9.992 9.989 9.991 0.995 0.996 0.994 4.995 4.993 4.996 9.995 9.998 9.996

0.999 1.001 0.999 4.992 4.993 4.991 9.992 9.989 9.991 0.995 0.996 0.994 4.995 4.993 4.996 9.995 9.998 9.996

10.038 10.038 10.038

10 11 12 13 14 15 16 17 18

1.009 1.009 1.009 5.019 5.019 5.019

10.038 10.038 10.038

Standards Used:

Description

I.D.

Range

Units Accuracy

Cal. Due

Traceability #

WEIGHT SET 1 CLASS F PRIMARY STANDARD

10040 0.016 - 50

lb in

0.01% 0.05%

27-Sep-2025 31-Dec-2026

45435-081061

MOUNTZ W2.0 PRIMARY STANDARD RADIUS WHEEL 43371 2

731133

Work Performed By: T. HOLLEY TECHNICIAN

Reviewed By: CALIBRATION QC A. RHODES

Calibrations are performed using standards traceable through NIST or other National Metrology Institute to the International System of Units (SI). Calibration system complies with ANSI/NCSL Z540-1 and ISO17025. This information applies only to the instrument identified above and may not be reproduced, except in full, without prior written consent. Reported uncertainties are expressed as expanded values at approximately 95% confidence level using coverage factor K=2. There is no implied warranty that instrument will maintain its specified tolerance during the calibration interval. Decision Rule 1 is employed (measurement uncertainty is not taken into account) when making conformity statements unless otherwise specified .

Page 1 of 1

Calibration Certificate Certificate No. 746737 216303 Mountz Cal ID:

www.mountztorque.com

# AC-1346

Calibration Performed By: MOUNTZ ALABAMA 19051 UNDERWOOD RD. FOLEY, AL 36535 USA (408) 850-9978

Issued To:

19051 UNDERWOOD RD. FOLEY, AL 36535 USA MOUNTZ ALABAMA

Sales Order Number: SO268939

Equipment Tested:

Cal. Date:

12 MONTHS 08-May-2026 08-May-2025

Description:

MOUNTZ FG-20I TORQUE SCREWDRIVER PRESET LIMITING

Cal. Interval:

Arrival Condition: Cal. Due Date:

Serial Number: Model Number:

0232128-3112

MEETS MFG SPEC

FG-20I

F Service Performed: CAL TO MFG SPEC / Temp. / RH%: 70 40

Range:

2 - 20

Unit of Measure: lbf.in

Calibration Result:

1.9% PASS

Accuracy:

6% OF SETTING

Uncertainty: Procedure:

3WI-7.1-001

Calibration Notes:

Measurement Data:

Seq.

Description Preset Reading 1 Preset Reading 2 Preset Reading 3 Preset Reading 4 Preset Reading 5

Nominal

Tolerance -

Tolerance +

As Found As Left

1 2 3 4 5

10.000 10.000 10.000 10.000 10.000

9.400 9.400 9.400 9.400 9.400

10.600 10.600 10.600 10.600 10.600

9.946

9.946

10.322 10.370 10.388 10.333

10.322 10.370 10.388 10.333

Standards Used:

Traceability #

Description

I.D.

Range

Units Accuracy

Cal. Due

MOUNTZ BMX100I TORQUE TRANSDUCER FIXED

193310 10 - 100

lbf.in

0.5%

03-Jun-2025

728853

Work Performed By: R. COUCH TECHNICIAN

Reviewed By: CALIBRATION QC T. HOLLEY

Calibrations are performed using standards traceable through NIST or other National Metrology Institute to the International System of Units (SI). Calibration system complies with ANSI/NCSL Z540-1 and ISO17025. This information applies only to the instrument identified above and may not be reproduced, except in full, without prior written consent. Reported uncertainties are expressed as expanded values at approximately 95% confidence level using coverage factor K=2. There is no implied warranty that instrument will maintain its specified tolerance during the calibration interval. Decision Rule 1 is employed (measurement uncertainty is not taken into account) when making conformity statements unless otherwise specified .

Page 1 of 1

Calibration Certificate Certificate No. 746570 216304 Mountz Cal ID:

www.mountztorque.com

# AC-1346

Calibration Performed By: MOUNTZ ALABAMA 19051 UNDERWOOD RD. FOLEY, AL 36535 USA (408) 850-9978

Issued To:

19051 UNDERWOOD RD. FOLEY, AL 36535 USA MOUNTZ ALABAMA

Sales Order Number: SO268939

Equipment Tested:

Cal. Date:

12 MONTHS 08-May-2026 08-May-2025

Description:

MOUNTZ FGC-125 TORQUE WRENCH PRESET CAM-OVER

Cal. Interval:

Arrival Condition: Cal. Due Date:

Serial Number: Model Number:

1012203-007

MEETS MFG SPEC

FGC-125 18.5 - 92.2

F Service Performed: CAL TO MFG SPEC / Temp. / RH%: 70 40

Range:

Unit of Measure: lbf.ft

Calibration Result:

0.92% PASS

Accuracy:

4% OF SETTING

Uncertainty: Procedure:

3WI-7.1-001

Calibration Notes:

Measurement Data:

Seq.

Description Preset Reading 1 Preset Reading 2 Preset Reading 3 Preset Reading 4 Preset Reading 5

Nominal

Tolerance -

Tolerance +

As Found As Left

1 2 3 4 5

75.000 75.000 75.000 75.000 75.000

72.000 72.000 72.000 72.000 72.000

78.000 78.000 78.000 78.000 78.000

75.680 76.603 76.369 76.585 76.521

75.680 76.603 76.369 76.585 76.521

Standards Used:

Traceability #

Description

I.D.

Range

Units Accuracy

Cal. Due

MOUNTZ BMX100F TORQUE TRANSDUCER FIXED 174516 10 - 100

lbf.ft

03-Jun-2025

728848

Work Performed By: R. COUCH TECHNICIAN

Reviewed By: CALIBRATION QC T. HOLLEY

Calibrations are performed using standards traceable through NIST or other National Metrology Institute to the International System of Units (SI). Calibration system complies with ANSI/NCSL Z540-1 and ISO17025. This information applies only to the instrument identified above and may not be reproduced, except in full, without prior written consent. Reported uncertainties are expressed as expanded values at approximately 95% confidence level using coverage factor K=2. There is no implied warranty that instrument will maintain its specified tolerance during the calibration interval. Decision Rule 1 is employed (measurement uncertainty is not taken into account) when making conformity statements unless otherwise specified .

Page 1 of 1

Calibration Certificate Certificate No. 736020 68135 Mountz Cal ID:

www.mountztorque.com

# AC-1346

Calibration Performed By: MOUNTZ ALABAMA 19051 UNDERWOOD RD. FOLEY, AL 36535 USA (408) 850-9978

Issued To:

19051 UNDERWOOD RD. FOLEY, AL 36535 USA MOUNTZ ALABAMA

Sales Order Number: SO265463

Equipment Tested:

Cal. Date:

6 MONTHS 04-Aug-2025 04-Feb-2025

Description:

MOUNTZ LTT-50I TORQUE ANALYZER ELECTRONIC

Cal. Interval:

Arrival Condition: Cal. Due Date:

Serial Number: Model Number:

LTT1106023

MEETS MFG SPEC

LTT-50I

F Service Performed: CAL TO MFG SPEC / Temp. / RH%: 70 42

Range:

5 - 50

Unit of Measure: lbf.in

Calibration Result:

0.13% PASS

Accuracy:

1% of reading from 10-20%FS 0.5% of reading from 20-100%FS

Uncertainty: Procedure:

3WI-7.1-003

Calibration Notes:

Calibrated using version 2.5.2.5. rev 1.23. Gain checked. A/D count reading: +20

Measurement Data:

Seq.

Description

Nominal

Tolerance -

Tolerance +

As Found As Left

1 2 3 4 5 6 7 8 9

CW 10% Reading 1 CW 10% Reading 2 CW 10% Reading 3 CW 50% Reading 1 CW 50% Reading 2 CW 50% Reading 3 CW 100% Reading 1 CW 100% Reading 2 CW 100% Reading 3 CCW 10% Reading 1 CCW 10% Reading 2 CCW 10% Reading 3 CCW 50% Reading 1 CCW 50% Reading 2 CCW 50% Reading 3 CCW 100% Reading 1 CCW 100% Reading 2 CCW 100% Reading 3

4.9941 4.9941 4.9941 24.9704 24.9704 24.9704 49.9408 49.9408 49.9408 4.9941 4.9941 4.9941 24.9704 24.9704 24.9704 49.9408 49.9408 49.9408

4.9441 4.9441 4.9441 24.7207 24.7207 24.7207 49.4414 49.4414 49.4414 4.9441 4.9441 4.9441 24.7207 24.7207 24.7207 49.4414 49.4414 49.4414

5.0440 5.0440 5.0440 25.2201 25.2201 25.2201 50.4402 50.4402 50.4402 5.0440 5.0440 5.0440 25.2201 25.2201 25.2201 50.4402 50.4402 50.4402

5.0110 5.0028 5.0055 25.0242 25.0269 25.0296 50.0167 50.0194 50.0249 4.9669 4.9641 4.9669 25.0112 25.0139 25.0084 50.0168 50.0389 50.0244

5.0110 5.0028 5.0055 25.0242 25.0269 25.0296 50.0167 50.0194 50.0249 4.9669 4.9641 4.9669 25.0112 25.0139 25.0084 50.0168 50.0389 50.0244

10 11 12 13 14 15 16 17 18

Standards Used:

Description

I.D.

Range

Units Accuracy

Cal. Due

Traceability #

WEIGHT SET 2 CLASS F PRIMARY STANDARD

10041 .016 - 50

lb in

0.01% 0.05%

03-Apr-2025 24-023 24-024 24-025

MOUNTZ W4.0 PRIMARY STANDARD RADIUS WHEEL 73354 4

08-Oct-2026

724448

Work Performed By: T. HOLLEY TECHNICIAN

Reviewed By: CALIBRATION QC A. RHODES

Calibrations are performed using standards traceable through NIST or other National Metrology Institute to the International System of Units (SI). Calibration system complies with ANSI/NCSL Z540-1 and ISO17025. This information applies only to the instrument identified above and may not be reproduced, except in full, without prior written consent. Reported uncertainties are expressed as expanded values at approximately 95% confidence level using coverage factor K=2. There is no implied warranty that instrument will maintain its specified tolerance during the calibration interval. Decision Rule 1 is employed (measurement uncertainty is not taken into account) when making conformity statements unless otherwise specified .

Page 1 of 1

Calibration Certificate Certificate No. 746723 30243 Mountz Cal ID:

www.mountztorque.com

# AC-1346

Calibration Performed By: MOUNTZ ALABAMA 19051 UNDERWOOD RD. FOLEY, AL 36535 USA (408) 850-9978

Issued To:

19051 UNDERWOOD RD. FOLEY, AL 36535 USA MOUNTZ ALABAMA

Sales Order Number: SO268939

Equipment Tested:

Cal. Date:

6 MONTHS 08-Nov-2025 08-May-2025

Description:

MOUNTZ LTT-250I TORQUE ANALYZER ELECTRONIC

Cal. Interval:

Arrival Condition: Cal. Due Date:

Serial Number: Model Number:

LTT0711024

OUT OF MFG SPEC

LTT-250I 25 - 250

F Service Performed: CAL TO MFG SPEC / Temp. / RH%: 70 40

Range:

Unit of Measure: lbf.in

Calibration Result:

ADJ.

Accuracy:

1% of reading from 10-20%FS 0.5% of reading from 20-100%FS

Uncertainty: Procedure:

0.13%

3WI-7.1-003

Calibration Notes:

Calibrated using version 2.5.2.5. rev 1.23. Gain checked. A/D +155

Measurement Data:

Seq.

Description

Nominal

Tolerance -

Tolerance +

As Found As Left

1 2 3 4 5 6 7 8 9

CW 10% Reading 1 CW 10% Reading 2 CW 10% Reading 3 CW 50% Reading 1 CW 50% Reading 2 CW 50% Reading 3 CW 100% Reading 1 CW 100% Reading 2 CW 100% Reading 3 CCW 10% Reading 1 CCW 10% Reading 2 CCW 10% Reading 3 CCW 50% Reading 1 CCW 50% Reading 2 CCW 50% Reading 3 CCW 100% Reading 1 CCW 100% Reading 2 CCW 100% Reading 3

24.970 24.970 24.970 124.852 124.852 124.852 249.704 249.704 249.704 24.970 24.970 24.970 124.852 124.852 124.852 249.704 249.704 249.704

24.721 24.721 24.721 124.228 124.228 124.228 248.455 248.455 248.455 24.721 24.721 24.721 124.228 124.228 124.228 248.455 248.455 248.455

25.220 25.220 25.220 125.476 125.476 125.476 250.953 250.953 250.953 25.220 25.220 25.220 125.476 125.476 125.476 250.953 250.953 250.953

A A A

24.597 24.624 24.610 124.568 124.675 124.595 249.732 249.679 249.719 25.308 25.295 25.281 125.283 125.243 125.203 250.296 250.122 250.162

25.081 25.094 25.107 125.001 124.987 124.974 250.176 249.961 249.988 25.121 25.134 25.147 125.036 125.063 125.009 249.907 250.068 250.041

10 11 12 13 14 15 16 17 18

A A A

Standards Used:

Description

I.D.

Range

Units Accuracy

Cal. Due

Traceability #

WEIGHT SET 1 CLASS F PRIMARY STANDARD

10040 0.016 - 50

lb in

0.01% 0.05%

27-Sep-2025 08-Oct-2026

45435-081061

MOUNTZ W4.0 PRIMARY STANDARD RADIUS WHEEL 73354 4

724448

Work Performed By: R. COUCH TECHNICIAN

Reviewed By: CALIBRATION QC T. HOLLEY

Calibrations are performed using standards traceable through NIST or other National Metrology Institute to the International System of Units (SI). Calibration system complies with ANSI/NCSL Z540-1 and ISO17025. This information applies only to the instrument identified above and may not be reproduced, except in full, without prior written consent. Reported uncertainties are expressed as expanded values at approximately 95% confidence level using coverage factor K=2. There is no implied warranty that instrument will maintain its specified tolerance during the calibration interval. Decision Rule 1 is employed (measurement uncertainty is not taken into account) when making conformity statements unless otherwise specified .

Page 1 of 1