Calibration Certificate Certificate No. 740301 169841 Mountz Cal ID:
www.mountztorque.com
# AC-1346
Calibration Performed By: MOUNTZ ALABAMA 19051 UNDERWOOD RD. FOLEY, AL 36535 USA (408) 850-9978
Issued To:
19051 UNDERWOOD RD. FOLEY, AL 36535 USA MOUNTZ ALABAMA
Sales Order Number: SO266691
Equipment Tested:
Cal. Date:
12 MONTHS 13-Mar-2026 13-Mar-2025
Description:
MOUNTZ BMX2500F TORQUE TRANSDUCER FIXED
Cal. Interval:
Arrival Condition: Cal. Due Date:
Serial Number: Model Number:
BMX1907047 BMX2500F 250 - 2,500
MEETS MFG SPEC
F Service Performed: CAL TO MFG SPEC / Temp. / RH%: 70 42
Range:
Unit of Measure: lbf.ft
Calibration Result:
0.15% PASS
Accuracy:
1% OF READING FROM 10 - 20% FS 0.5% OF READING FROM 20 - 100% FS
Uncertainty: Procedure:
3WI-7.1-004
Calibration Notes:
Calibrated using PTT s/n: PTT0607118 version 2.5.2.5. rev 1.26. Gain checked. A/D count reading: +155
Measurement Data:
Seq.
Description
Nominal
Tolerance -
Tolerance +
As Found As Left
1 2 3 4 5 6 7 8 9
CW 10% Reading 1 CW 10% Reading 2 CW 10% Reading 3 CW 50% Reading 1 CW 50% Reading 2 CW 50% Reading 3 CW 100% Reading 1 CW 100% Reading 2 CW 100% Reading 3 CCW 10% Reading 1 CCW 10% Reading 2 CCW 10% Reading 3 CCW 50% Reading 1 CCW 50% Reading 2 CCW 50% Reading 3 CCW 100% Reading 1 CCW 100% Reading 2 CCW 100% Reading 3
250.00 250.00 250.00 1250.00 1250.00 1250.00 2500.00 2500.00 2500.00 250.00 250.00 250.00 1250.00 1250.00 1250.00 2500.00 2500.00 2500.00
247.50 247.50 247.50 1243.75 1243.75 1243.75 2487.50 2487.50 2487.50 247.50 247.50 247.50 1243.75 1243.75 1243.75 2487.50 2487.50 2487.50
252.50 252.50 252.50 1256.25 1256.25 1256.25 2512.50 2512.50 2512.50 252.50 252.50 252.50 1256.25 1256.25 1256.25 2512.50 2512.50 2512.50
250.87 250.61 250.78 1247.38 1247.22 1247.10 2498.71 2498.66 2498.54 250.86 250.77 250.68 1249.63 1249.34 1249.57 2497.33 2497.11 2497.49
250.87 250.61 250.78 1247.38 1247.22 1247.10 2498.71 2498.66 2498.54 250.86 250.77 250.68 1249.63 1249.34 1249.57 2497.33 2497.11 2497.49
10 11 12 13 14 15 16 17 18
Standards Used:
Description
I.D.
Range
Units Accuracy
Cal. Due
Traceability #
1111HL-10K TORQUE TRANSDUCER LOAD CELL
10001 100 - 10000
lbf.ft
0.02%
25-Mar-2025 22-Jul-2026
S000003429
HP 34401-A MULTIMETER DIGITAL
73677 0 - 99
mV/V 0.0015%
19012428
Work Performed By: T. HOLLEY TECHNICIAN
Reviewed By: CALIBRATION QC A. RHODES
Calibrations are performed using standards traceable through NIST or other National Metrology Institute to the International System of Units (SI). Calibration system complies with ANSI/NCSL Z540-1 and ISO17025. This information applies only to the instrument identified above and may not be reproduced, except in full, without prior written consent. Reported uncertainties are expressed as expanded values at approximately 95% confidence level using coverage factor K=2. There is no implied warranty that instrument will maintain its specified tolerance during the calibration interval. Decision Rule 1 is employed (measurement uncertainty is not taken into account) when making conformity statements unless otherwise specified .