Calibration Certificate Certificate No. 746673 131973 Mountz Cal ID:
www.mountztorque.com
# AC-1346
Calibration Performed By: MOUNTZ ALABAMA 19051 UNDERWOOD RD. FOLEY, AL 36535 USA (408) 850-9978
Issued To:
19051 UNDERWOOD RD. FOLEY, AL 36535 USA MOUNTZ ALABAMA
Sales Order Number: SO268939
Equipment Tested:
Cal. Date:
6 MONTHS 08-Nov-2025 08-May-2025
Description:
MOUNTZ BMX20Z TORQUE TRANSDUCER FIXED
Cal. Interval:
Arrival Condition: Cal. Due Date:
Serial Number: Model Number:
BMX1211033
OUT OF MFG SPEC
BMX20Z
F Service Performed: CAL TO MFG SPEC / Temp. / RH%: 70 40
Range:
2 - 20
Unit of Measure: ozf.in
Calibration Result:
ADJ.
Accuracy:
1% OF READING FROM 10 - 20% FS 0.5% OF READING FROM 20 - 100% FS
Uncertainty: Procedure:
0.13%
3WI-7.1-003
Calibration Notes:
Calibrated using PTT s/n: 2206004 version 2.5.2.5. rev 1.26. Gain checked. A/D -102
Measurement Data:
Seq.
Description
Nominal
Tolerance -
Tolerance +
As Found As Left
1 2 3 4 5 6 7 8 9
CW 10% Reading 1 CW 10% Reading 2 CW 10% Reading 3 CW 50% Reading 1 CW 50% Reading 2 CW 50% Reading 3 CW 100% Reading 1 CW 100% Reading 2 CW 100% Reading 3 CCW 10% Reading 1 CCW 10% Reading 2 CCW 10% Reading 3 CCW 50% Reading 1 CCW 50% Reading 2 CCW 50% Reading 3 CCW 100% Reading 1 CCW 100% Reading 2 CCW 100% Reading 3
1.9976 1.9976 1.9976 9.9882 9.9882 9.9882 19.9763 19.9763 19.9763 1.9976 1.9976 1.9976 9.9882 9.9882 9.9882 19.9763 19.9763 19.9763
1.9777 1.9777 1.9777 9.9382 9.9382 9.9382 19.8764 19.8764 19.8764 1.9777 1.9777 1.9777 9.9382 9.9382 9.9382 19.8764 19.8764 19.8764
2.0176 2.0176 2.0176 10.0381 10.0381 10.0381 20.0762 20.0762 20.0762 2.0176 2.0176 2.0176 10.0381 10.0381 10.0381 20.0762 20.0762 20.0762
A A A A A A A A A A A A A
2.2025 2.2076 2.2054 10.2122 10.2151 10.2166 20.4274 20.4432 20.4497 1.6166 1.6173 1.6188 9.9322 9.9416 9.9395 19.9812 19.9950 19.9928
1.9812 1.9814 1.9835 9.9452 9.9473 9.9489 20.0118 20.0216 20.0230 1.9809 1.9816 1.9823 9.9639 9.9667 9.9681 20.0072 20.0016 20.0009
10 11 12 13 14 15 16 17 18
Standards Used:
Description
I.D.
Range
Units Accuracy
Cal. Due
Traceability #
WEIGHT SET 1 CLASS F PRIMARY STANDARD
10040 0.016 - 50
lb in
0.01%
27-Sep-2025 15-Feb-2026
45435-081061
MOUNTZ W1.0 PRIMARY STANDARD RADIUS WHEEL 250335 1
0.1%
692580
Work Performed By: R. COUCH TECHNICIAN
Reviewed By: CALIBRATION QC T. HOLLEY
Calibrations are performed using standards traceable through NIST or other National Metrology Institute to the International System of Units (SI). Calibration system complies with ANSI/NCSL Z540-1 and ISO17025. This information applies only to the instrument identified above and may not be reproduced, except in full, without prior written consent. Reported uncertainties are expressed as expanded values at approximately 95% confidence level using coverage factor K=2. There is no implied warranty that instrument will maintain its specified tolerance during the calibration interval. Decision Rule 1 is employed (measurement uncertainty is not taken into account) when making conformity statements unless otherwise specified .
Page 1 of 1
Made with FlippingBook Annual report maker