Mountz Sample Calibration Certifications

Calibration Certificate Certificate No. 746737 216303 Mountz Cal ID:

www.mountztorque.com

# AC-1346

Calibration Performed By: MOUNTZ ALABAMA 19051 UNDERWOOD RD. FOLEY, AL 36535 USA (408) 850-9978

Issued To:

19051 UNDERWOOD RD. FOLEY, AL 36535 USA MOUNTZ ALABAMA

Sales Order Number: SO268939

Equipment Tested:

Cal. Date:

12 MONTHS 08-May-2026 08-May-2025

Description:

MOUNTZ FG-20I TORQUE SCREWDRIVER PRESET LIMITING

Cal. Interval:

Arrival Condition: Cal. Due Date:

Serial Number: Model Number:

0232128-3112

MEETS MFG SPEC

FG-20I

F Service Performed: CAL TO MFG SPEC / Temp. / RH%: 70 40

Range:

2 - 20

Unit of Measure: lbf.in

Calibration Result:

1.9% PASS

Accuracy:

6% OF SETTING

Uncertainty: Procedure:

3WI-7.1-001

Calibration Notes:

Measurement Data:

Seq.

Description Preset Reading 1 Preset Reading 2 Preset Reading 3 Preset Reading 4 Preset Reading 5

Nominal

Tolerance -

Tolerance +

As Found As Left

1 2 3 4 5

10.000 10.000 10.000 10.000 10.000

9.400 9.400 9.400 9.400 9.400

10.600 10.600 10.600 10.600 10.600

9.946

9.946

10.322 10.370 10.388 10.333

10.322 10.370 10.388 10.333

Standards Used:

Traceability #

Description

I.D.

Range

Units Accuracy

Cal. Due

MOUNTZ BMX100I TORQUE TRANSDUCER FIXED

193310 10 - 100

lbf.in

0.5%

03-Jun-2025

728853

Work Performed By: R. COUCH TECHNICIAN

Reviewed By: CALIBRATION QC T. HOLLEY

Calibrations are performed using standards traceable through NIST or other National Metrology Institute to the International System of Units (SI). Calibration system complies with ANSI/NCSL Z540-1 and ISO17025. This information applies only to the instrument identified above and may not be reproduced, except in full, without prior written consent. Reported uncertainties are expressed as expanded values at approximately 95% confidence level using coverage factor K=2. There is no implied warranty that instrument will maintain its specified tolerance during the calibration interval. Decision Rule 1 is employed (measurement uncertainty is not taken into account) when making conformity statements unless otherwise specified .

Page 1 of 1