Mountz Sample Calibration Certifications

Calibration Certificate Certificate No. 746570 216304 Mountz Cal ID:

www.mountztorque.com

# AC-1346

Calibration Performed By: MOUNTZ ALABAMA 19051 UNDERWOOD RD. FOLEY, AL 36535 USA (408) 850-9978

Issued To:

19051 UNDERWOOD RD. FOLEY, AL 36535 USA MOUNTZ ALABAMA

Sales Order Number: SO268939

Equipment Tested:

Cal. Date:

12 MONTHS 08-May-2026 08-May-2025

Description:

MOUNTZ FGC-125 TORQUE WRENCH PRESET CAM-OVER

Cal. Interval:

Arrival Condition: Cal. Due Date:

Serial Number: Model Number:

1012203-007

MEETS MFG SPEC

FGC-125 18.5 - 92.2

F Service Performed: CAL TO MFG SPEC / Temp. / RH%: 70 40

Range:

Unit of Measure: lbf.ft

Calibration Result:

0.92% PASS

Accuracy:

4% OF SETTING

Uncertainty: Procedure:

3WI-7.1-001

Calibration Notes:

Measurement Data:

Seq.

Description Preset Reading 1 Preset Reading 2 Preset Reading 3 Preset Reading 4 Preset Reading 5

Nominal

Tolerance -

Tolerance +

As Found As Left

1 2 3 4 5

75.000 75.000 75.000 75.000 75.000

72.000 72.000 72.000 72.000 72.000

78.000 78.000 78.000 78.000 78.000

75.680 76.603 76.369 76.585 76.521

75.680 76.603 76.369 76.585 76.521

Standards Used:

Traceability #

Description

I.D.

Range

Units Accuracy

Cal. Due

MOUNTZ BMX100F TORQUE TRANSDUCER FIXED 174516 10 - 100

lbf.ft

03-Jun-2025

728848

Work Performed By: R. COUCH TECHNICIAN

Reviewed By: CALIBRATION QC T. HOLLEY

Calibrations are performed using standards traceable through NIST or other National Metrology Institute to the International System of Units (SI). Calibration system complies with ANSI/NCSL Z540-1 and ISO17025. This information applies only to the instrument identified above and may not be reproduced, except in full, without prior written consent. Reported uncertainties are expressed as expanded values at approximately 95% confidence level using coverage factor K=2. There is no implied warranty that instrument will maintain its specified tolerance during the calibration interval. Decision Rule 1 is employed (measurement uncertainty is not taken into account) when making conformity statements unless otherwise specified .

Page 1 of 1