Calibration Certificate Certificate No. 736020 68135 Mountz Cal ID:
www.mountztorque.com
# AC-1346
Calibration Performed By: MOUNTZ ALABAMA 19051 UNDERWOOD RD. FOLEY, AL 36535 USA (408) 850-9978
Issued To:
19051 UNDERWOOD RD. FOLEY, AL 36535 USA MOUNTZ ALABAMA
Sales Order Number: SO265463
Equipment Tested:
Cal. Date:
6 MONTHS 04-Aug-2025 04-Feb-2025
Description:
MOUNTZ LTT-50I TORQUE ANALYZER ELECTRONIC
Cal. Interval:
Arrival Condition: Cal. Due Date:
Serial Number: Model Number:
LTT1106023
MEETS MFG SPEC
LTT-50I
F Service Performed: CAL TO MFG SPEC / Temp. / RH%: 70 42
Range:
5 - 50
Unit of Measure: lbf.in
Calibration Result:
0.13% PASS
Accuracy:
1% of reading from 10-20%FS 0.5% of reading from 20-100%FS
Uncertainty: Procedure:
3WI-7.1-003
Calibration Notes:
Calibrated using version 2.5.2.5. rev 1.23. Gain checked. A/D count reading: +20
Measurement Data:
Seq.
Description
Nominal
Tolerance -
Tolerance +
As Found As Left
1 2 3 4 5 6 7 8 9
CW 10% Reading 1 CW 10% Reading 2 CW 10% Reading 3 CW 50% Reading 1 CW 50% Reading 2 CW 50% Reading 3 CW 100% Reading 1 CW 100% Reading 2 CW 100% Reading 3 CCW 10% Reading 1 CCW 10% Reading 2 CCW 10% Reading 3 CCW 50% Reading 1 CCW 50% Reading 2 CCW 50% Reading 3 CCW 100% Reading 1 CCW 100% Reading 2 CCW 100% Reading 3
4.9941 4.9941 4.9941 24.9704 24.9704 24.9704 49.9408 49.9408 49.9408 4.9941 4.9941 4.9941 24.9704 24.9704 24.9704 49.9408 49.9408 49.9408
4.9441 4.9441 4.9441 24.7207 24.7207 24.7207 49.4414 49.4414 49.4414 4.9441 4.9441 4.9441 24.7207 24.7207 24.7207 49.4414 49.4414 49.4414
5.0440 5.0440 5.0440 25.2201 25.2201 25.2201 50.4402 50.4402 50.4402 5.0440 5.0440 5.0440 25.2201 25.2201 25.2201 50.4402 50.4402 50.4402
5.0110 5.0028 5.0055 25.0242 25.0269 25.0296 50.0167 50.0194 50.0249 4.9669 4.9641 4.9669 25.0112 25.0139 25.0084 50.0168 50.0389 50.0244
5.0110 5.0028 5.0055 25.0242 25.0269 25.0296 50.0167 50.0194 50.0249 4.9669 4.9641 4.9669 25.0112 25.0139 25.0084 50.0168 50.0389 50.0244
10 11 12 13 14 15 16 17 18
Standards Used:
Description
I.D.
Range
Units Accuracy
Cal. Due
Traceability #
WEIGHT SET 2 CLASS F PRIMARY STANDARD
10041 .016 - 50
lb in
0.01% 0.05%
03-Apr-2025 24-023 24-024 24-025
MOUNTZ W4.0 PRIMARY STANDARD RADIUS WHEEL 73354 4
08-Oct-2026
724448
Work Performed By: T. HOLLEY TECHNICIAN
Reviewed By: CALIBRATION QC A. RHODES
Calibrations are performed using standards traceable through NIST or other National Metrology Institute to the International System of Units (SI). Calibration system complies with ANSI/NCSL Z540-1 and ISO17025. This information applies only to the instrument identified above and may not be reproduced, except in full, without prior written consent. Reported uncertainties are expressed as expanded values at approximately 95% confidence level using coverage factor K=2. There is no implied warranty that instrument will maintain its specified tolerance during the calibration interval. Decision Rule 1 is employed (measurement uncertainty is not taken into account) when making conformity statements unless otherwise specified .
Page 1 of 1