Press the Place Table button

Place table values

4 Modify the values of the place table to represent the values found in the detail block of your engineering drawing if required, then press the Done button to return to the Feature Detail screen. Measure toolbar > Done button 5 Navigate back to the Nominal Feature Detail screen if necessary, and enter the nominal values using one of the methods described earlier. Click the non-significant digits in each field to apply the Place Table values. Non-significant digits will become gray to indicate that the values have been applied. Press the Done button to conclude this process. Measure toolbar > Done button

Press non-significant digits

Enter nominal values

6 Press the Nominal button to go to the Tolerance Feature Detail screen. Tolerances from the Place Tolerance table will have been applied. 7 Press the Tolerance button to go to the Deviation Feature Detail screen and observe the results of the tolerance tests.