Video Inspection Software User’s Guide
M3 system menu > Settings > Contact Probes 2 Select the probe tip to be edited from the probe group. 3 Press the Edit button to display the Edit Contact Probe dialog. Changing the probe Tip Name To change the name: 1 Enter the new name into the Name field and press Done.
Assigning a Tip Type To assign a tip type: 1 Press the Tip Type drop-down box button, select the desired type and press Done. Available tip type assignments include: • Spherical • Disk • Cylindrical
Designating the Probe as a Reference Any probe in the probe group can be designated as the reference.
CAUTION When the reference probe tip is changed, the current calibrations of all other tips in the group become invalidated and after the calibration of the new reference tip, new calibrations of all other probes will need to be performed.
To designate a probe as the reference, press the Reference drop-down box, select Yes and press Done.
Changing the Current Probe Tip
To change the current probe tip: 1 Press the Contact Probes Settings screen button to access the Contact Probe Settings screen. M3 system menu > Settings > Contact Probes 2 Select the desired probe in the probe group and press the Set Current button. The selected probe will now be shown as current in the probe group.
45