Composite Brush Handle

Processing: - Clean, disinfect and sterilize in a completely disassembled state.

Container for Sterilization and Accessories

Processing: - For reprocessing, the lid of the Container and the filter holding devices in the base and the lid must be removed. - If single use paper filters have been used, they must be removed before reprocessing. Indica- tors must be removed from the label holding device. Cleaning / Disinfection: - For the cleaning and disinfection of anodized Aluminium Sterilization Containers only deter- gents and disinfectants can be used which are approved for this material. - For the reprocessing in a Washer-Disinfector unit the components of the Container have to be placed securely in the washing baskets. Spraying nozzles and arms should not be blocked. Do not use acid neutralizers for the reprocessing of Aluminium Containers. - Container made of Aluminium cannot be cleaned or disinfected in an Ultrasonic Cleaner Unit. - Container-Cassettes may be cleaned and disinfected with all procedures recommended for IMS-cassettes. - Permanent Filters of Teflon may be cleaned and disinfected in a Washer-Disinfector unit. - Wrapping Drapes for aseptic removal must be cleaned with the standard procedures for clinical or dental office textiles, before reuse. Do not starch wrapping drapes

Sterilization: -

Sterilization Containers made of anodized Aluminium have been developed especially for ster- ilization in Steam Sterilizers with pre-vacuum, fractioned vacuum or fractioned flow processes. Hu-Friedy Sterilization Containers cannot be used for other sterilization methods.

Maintenance: - The surface of Aluminium Containers is very sensitive in respect to mechanical impact. For this reason, do not use metal brushes or scouring agents. - For the removal of stains, residues of inscriptions or adhesive tapes only a commercial cleaner for anodized Aluminium may be used (no benzine or acetone!) After such treatment the Containers must be cleaned once more.

Crown Remover (CRL, CRU)

Processing: - Clean, disinfect and sterilize in a completely disassembled state.

Cleaning / Disinfection: -

Do not disinfect with phenols or iodophors.

Sterilization: -

Do not sterilize with dry heat.

IMPLACARE

Sterilization: - IMPLACARE disposable resin tips can be steam sterilized prior to use. - They are intended for one use only! - Clean, disinfect and sterilize in a completely disassembled state.

Mallet

Processing: - Clean, disinfect and sterilize in a completely disassembled state.

Mouth Gags (MGA, MGC, MGI)

Processing: -

When using a cassette system for cleaning/sterilization, the opening where the nylon tubing slips over the instrument tip must not be covered so as to allow the tips to properly drain. Clean, disinfect and sterilize in a completely disassembled state.