HF-GB-17664, Rev. D | Page 8 / 8 Status: 07/2023

Scaler marked with Color Coding Rings (IMS-1280L, IMS-1286L, IMS-1281, IMS-1287, IMS-12810, IMS-1287L, IMS- 12810L, IMS-1288, IMS-12811, IMS- 1288L, IMS-12811L, IMS-1289, IMS-1281L, IMS- 1289L, IMS-1282, IMS-1282L, IMS-1283, IMS-1283L, IMS-1284, IMSö1284L, IMS- 1234, IMS-1285, IMS-1285L, IMS-1280, IMS-1286)

Processing: - For reprocessing, attached Color Coding Rings do not have to be removed

Validation reports: -

Project Number:00418-1 Examination of an Automated Cleaning Process using quantitative Detection of Protein and Hemoglobin and the Radionuclide Method 07-Jun-2018 - Project Number: 00418-2 Examination of a Manual Cleaning Process using quantitative Detection of Protein and Hemoglobin and the Radionuclide Method 07-Jun-2018 - Project Number: 25517-1 Validation of a Sterilization Process using Steam Sterilization in Pre-vacuum Mode Method MD 4.0: Sterilization validation of medical products with moist heat 12-Dec-2018 - Project Number: 25517-2 Validation of a Sterilization Process using Steam Sterilization in Pre-vacuum Mode Method MD 4.0: Sterilization validation of medical products with moist heat 12-Dec-2018

Syringes

Processing: - Completely disassemble including unscrewing of the cylinder.

Ultrasonic Inserts, Magnetostrictive

Processing: - Ultrasonic cleaning and disinfection as well as steam sterilization can be affected in suitable Hu-Friedy IMS-Cassettes.

Cleaning / Disinfection: -

For automated cleaning and disinfection in a Washer-Disinfector unit connecting rinsing adapters must be used, if the inserts are processed inside a cassette system. Otherwise open tray systems for the automated cleaning and disinfection or alternatively the manual cleaning and disinfection procedure are recommended.

Sterilization: - For sterilization use steam sterilization only. Do not expose to phenols or iodophors. - Do not use dry heat sterilization, or heat above 135 °C (275 F).

Ultrasonic Inserts, Piezo with Guardian

Processing: - Piezo Ultrasonic Inserts remain in the Guardian during the complete reprocessing cycle, also if reprocessed in cassettes. - Ultrasonic cleaning and disinfection as well as steam sterilization can be affected in suitable Hu-Friedy IMS-Cassettes.

Sterilization: -

For sterilization use steam sterilization only. Do not expose to phenols or iodophors.

-

- Do not use dry heat sterilization, or heat above 135 °C (275 F).

Ultrasonic Piezo Handpiece

Sterilization: - The Piezo handpiece can be steam sterilized with all types of Steam Sterilizers at 134°C/ 15 minutes. Other sterilization parameters are not permitted.

Free Call: 00800 48 37 43 39 | Free Fax: 00800 48 37 43 40 E-Mail: info@hu-friedy.eu | Website: HuFriedyGroup.eu HuFriedyGroupEurope

Hu-Friedy Mfg. Co., LLC. • European Headquarters • Lyoner Str. 9 • 60528 Frankfurt am Main, Germany • HuFriedyGroup.eu All company and product names are trademarks of Hu-Friedy Mfg. Co. LLC., its affiliates or related companies, unless otherwise noted. ©2023 Hu-Friedy Mfg. Co., LLC. All rights reserved. HF-GB-17664/0723