The EC-Series DC Torque Control System Calibration Procedure explains how to verify, calibrate, and adjust torque output for optimal fastening performance. Learn the four key calibration scenarios, including factory calibration, preset torque certification, on-site verification, and torque transfer to the application, to ensure accurate, repeatable results and process confidence. Designed for a cleanroom environment, the FG-CR cleanroom screwdriver is manufactured using a high-grade, low-outgassing powder coat. The tools feature dynamic wear-resistant internal seals to prevent grease leakage and minimize airborne contamination.
CERTIFICATE OF ACCREDITATION
The ANSI National Accreditation Board
Hereby attests that
Mountz, Inc. 19051 Underwood Rd. Foley, AL 36535
Fulfills the requirements of
ISO/IEC 17025:2017
In the field of
CALIBRATION
This certificate is valid only when accompanied by a current scope of accreditation document. The current scope of accreditation can be verified at www.anab.org.
______________________________ Jason Stine, Vice President
Expiry Date: 29 October 2028 Certificate Number: AC-1346.01
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017. This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality management system (refer to joint ISO-ILAC-IAF Communiqué dated April 2017).
SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017
Mountz, Inc. 19051 Underwood Rd. Foley, AL 36535 Ben Slaughter 251-336-3177
CALIBRATION
ISO/IEC 17025 Accreditation Granted: 30 September 2026
Certificate Number: AC-1346.01
Certificate Expiry Date: 29 October 2028
Mass and Mass Related
Reference Standard, Method, and/or Equipment Comparison to Lever Arms/Wheels, Dead Weights Comparison to Reference Load Cells
Expanded Uncertainty of Measurement (+/-)
Parameter/Equipment
Range
Torque Transducer
1 ozf·in to 1 000 lbf·ft
0.13 % of reading
Torque Transducer
(100 to 20 000) lbf·ft
0.15 % of reading
(1 to 160) ozf·in (10 to 750) lbf·in (50 to 2 500) lbf·ft (100 to 10 000) lbf·ft (1 to 160) ozf·in (10 to 750) lbf·in (50 to 500) lbf·ft (75 to 7 500) lbf·ft
2.75 % of reading 1.9 % of reading 0.92 % of reading 0.71 % of reading 3.5 % of reading 2.2 % of reading 1.04 % of reading 0.89 % of reading
Comparison to Digital Torque Tester
Torque Hand Tools 1
Comparison to Digital Torque Tester
Torque Power Tools 1
Calibration and Measurement Capability (CMC) is expressed in terms of the measurement parameter, measurement range, expanded uncertainty of measurement and reference standard, method, and/or equipment. The expanded uncertainty of measurement is expressed as the standard uncertainty of the measurement multiplied by a coverage factor of 2 ( k =2), corresponding to a confidence level of approximately 95%. Notes: 1. On-site calibration service is available for this parameter, since on-site conditions are typically more variable than those in the laboratory, larger measurement uncertainties are expected on-site than what is reported on the accredited scope. 2. Unless otherwise specified, the calibration method/procedure utilized by the client was developed internally.
This Scope of Accreditation, version 010, was last updated on: 30 September 2026 and is valid only when accompanied by the Certificate.
Page 1 of 1