SenSafe ® Iodine Test Strips
Dental Water Management
Testing Strip Procedure 1. Perform a two-minute flush of all dental unit waterlines. 2. Remove the syringe tip from air water syringe. 3. Remove one test strip from bottle and reseal bottle to maintain freshness of other test strips. 4. Take a 20mL water sample by tilting the cup (a plastic dental cup is acceptable) and allowing the water to run continuously down the interior side of the cup to avoid turbulence of the water. 5. Follow Method A as noted on the test strip bottle: a D ip one test strip for 10 seconds in the water sample using a constant, gentle back-and-forth motion. b Remove the strip from the water and wait 30 seconds. c Match color appearing in the small window of the test strip with the associated Method A color chart with ppm levels from 0.0 ppm – 5.0 ppm. For better color matching, fold the dipped strip in half, so that the window side of strip is facing out (see image below).
CORRECT:
AIR WATER SYRINGE WITH SYRINGE TIP REMOVED
TILTED CUP
ALLOW WATER TO RUN DOWN THE INTERIOR OF THE CUP
INCORRECT:
DO NOT SPRAY WATER FOR THE SAMPLE INTO AN UPRIGHT CUP
REMOVE THE AIR WATER SYRINGE TIP
Unfolded Strip
Folded strip with window facing out
Learn More at HuFriedyGroup.eu Free Call: 00800 48 37 43 39 | Free Fax: 00800 48 37 43 40 E-Mail: info@hu-friedy.eu | Website: HuFriedyGroup.eu HuFriedyGroupEurope
Hu-Friedy Mfg. Co., LLC. • European Headquarters • Lyoner Str. 9 • 60528 Frankfurt am Main, Germany • HuFriedyGroup.eu All company and product names are trademarks of Hu-Friedy Mfg. Co. LLC., its affiliates or related companies, unless otherwise noted. ©2024 Hu-Friedy Mfg. Co., LLC. All rights reserved. HFL-730GB/0524
Made with FlippingBook Annual report