HuFriedyGroup | SenSafe Iodine Test Strips EN

SenSafe ® Iodine Test Strips

Dental Water Management

Testing Strip Procedure 1. Perform a two-minute flush of all dental unit waterlines. 2. Remove the syringe tip from air water syringe. 3. Remove one test strip from bottle and reseal bottle to maintain freshness of other test strips. 4. Take a 20mL water sample by tilting the cup (a plastic dental cup is acceptable) and allowing the water to run continuously down the interior side of the cup to avoid turbulence of the water. 5. Follow Method A as noted on the test strip bottle:

CORRECT:

AIR WATER SYRINGE WITH SYRINGE TIP REMOVED

TILTED CUP

ALLOW WATER TO RUN DOWN THE INTERIOR OF THE CUP

a. Dip one test strip for 10 seconds in the water sample using a constant, gentle back-and-forth motion. a. Remove the strip from the water and wait 30 seconds.

INCORRECT:

DO NOT SPRAY WATER FOR THE SAMPLE INTO AN UPRIGHT CUP

a. Match color appearing in the small window of the test strip with the associated Method A color chart with ppm levels from 0.0 ppm – 5.0 ppm. For better color matching, fold the dipped strip in half, so that the window side of strip is facing out (see image below).

REMOVE THE AIR WATER SYRINGE TIP

Unfolded Strip

Folded strip with window facing out

Free Call: 00800 48 37 43 39 | Free Fax: 00800 48 37 43 40 E-Mail: info@hu-friedy.eu | Website: HuFriedyGroup.eu @HuFriedyGroupEurope Learn More at HuFriedyGroup.eu

Hu-Friedy Mfg. Co., LLC. • European Headquarters • Lyoner Str. 9 • 60528 Frankfurt am Main, Germany • HuFriedyGroup.eu All company and product names are trademarks of Hu-Friedy Mfg. Co. LLC., its affiliates or related companies, unless otherwise noted. ©2022 Hu-Friedy Mfg. Co., LLC. All rights reserved. HFL-730GB/0822