WHAT IS SEFA? The Scientific Equipment and Furniture Associ- ation (SEFA) is a voluntary international trade association representing members of the labora- tory furniture, casework, fume hood and related equipment industry. The Association was found- ed to promote this rapidly expanding industry and to improve the quality, safety and timely com- pletion of laboratory facilities in accordance with customer requirements. SEFA members work together to establish, monitor and modify, as needed, industry-wide recommended practices in the areas of fume hoods, laboratory work surfaces, equipment and furniture installations. SEFA RECOMMENDED PRACTICES SEFA and its committees are active in the devel- opment and promotion of recommended prac- tices having domestic and international applica- tions. Recommended practices are developed by the association taking into account the work of other national standard-writing organizations. Liaison is also maintained with government agen- cies in the development of their specifications. SEFA’s recommended practices are developed in and for the public interest. These practices are designed to promote better understanding between designers, architects, manufacturers, purchasers, and end users to assist the purchas- er in selecting and specifying the proper product to meet the user's particular needs. LABORATORY FURNITURE - SEFA 8 This recommended practice is intended to pro- vide manufacturers, specifiers, and users tools for evaluating the safety, durability, and structural integrity of laboratory casework and compli- mentary items.
SEFA 8 CERTIFICATE OF PERFORMANCE INDEPENDENT TEST LAB
Diversified Woodcrafts, Inc. contracted with Bjorksten Research Laboratory, 2 Fen Oak Court, Madison, WI 53718 to test our laboratory casework as it conforms to the SEFA 8 stan- dards. The finish tests were performed by Chem- craft International, Inc., 1101 N. Bluemound Drive, Appleton, WI 54914. The summary results are shown below:
Test
Result
Test
Result
4.2 4.3 4.4 4.5 5.1 5.2 5.3 6.1 6.2
Pass Pass Pass Pass Pass Pass Pass Pass Pass
6.3 6.4 6.5 7.1 8.1 8.2 8.3 9.2
Pass Pass Pass Pass Pass Pass Pass Pass
DESCRIPTION OF TEST CABINET Diversified Woodcrafts, Inc. model number 108- 4822, 48" wide x 22" deep combination base cabinet with one (full opening) drawer and two doors. Cabinet was leveled and the drawers, doors, hinges, latches, etc. were inspected and all were operating properly. The wall cabinet was Diversified Woodcrafts, Inc. model number D03-4812, 30" high x 48" wide x 12" deep with 2 doors and 2 movable shelves.
DiversifiedCasework.com | info@DiversifiedCasework.com | 877.348.9663
SEFA Results
CABINET LOAD TEST
4.2.1 Purpose of Test The cabinet load test challenges the structural integrity and the load-bearing capability of the cabinet construction. The test demonstrates the ability of the cabinet to support heavy loads. 4.2.2 Test Procedure The cabinet top was loaded using 2,000 pounds of solid steel bars stacked five high, in eight equally spaced rows. After 10 minutes, the bars were removed from the cabinet. 4.2.3 Acceptance Level There are no signs of permanent damage. 4.3.1 Purpose of Test The purpose of the test is to challenge the functional characteristics of the cabinet when subjected to a concentrated load on the center of the cabinet top. 4.3.2 Test Procedure Four, 50 pound solid steel bars were placed on top of the cabinet, along the cabinet centerline. With the weight in place, the drawers and doors were operated and inspected. 4.3.3 Acceptance Level Doors and drawers operated normally. There is no permanent distortion to the front rail, cabinet joinery, doors, or drawers. CABINET CONCENTRATED LOAD TEST 4.4.1 Purpose of Test The test evaluates the structural integrity of the cabinet construction when subjected to torsional load. 4.4.2 Test Procedure The cabinet was supported on three, 3-1/4” x 3-1/2” x 4”H hardwood blocks, located under the two rear corners and one front corner. The cabinet was secured, diagonally from the unsupported corner with seven, 50 pound, solid steel bars to prevent overturning. Four, 50 pound, solid steel bares were placed on the unsupported corner of the cabinet and left in position for 15 minutes. The weights were then removed and the cabinet was placed on the floor in its normal, upright, position. The cabinet joinery was inspected and the cabinet was leveled and measured, diagonally, between the corners of the face and the back of the cabinet. 4.4.3 Acceptance Level When returned to normal position, the operation of the cabinet was normal and there was no evidence of permanent damage. The difference between the diagonal measurements does not exceed 1/8” front or back. 4.5.1 Purpose of Test The test demonstrated the ability of a cabinet to resist standing water. 4.5.2 Test Procedure The thickness of the material along the perimeter of the bottom of the cabinet was measured on 6” increments. The thickness was recorded and the arithmetic mean was calculated. The cabinet was then placed in a pan of water so that the bottom 2” of the cabinet was submerged. The cabinet remained in the water for 4 hours and then it was removed and immediately re-measured at the same locations that were measured ini- tially. The thickness at each location was recorded and the arithmetic mean was calculated. After the cabinet had dried, it was inspected for damage. 4.5.3 Acceptance Level The cabinet shows no signs of permanent deformation or deterioration and the average increase in thickness is less than four percent (3.5%). CABINET TORSION CABINET SUBMERSION TEST
DiversifiedCasework.com | info@DiversifiedCasework.com | 877.348.9663
SEFA Results
DOOR HINGE TEST
5.1.1 Purpose of Test The test demonstrates the durability of the door and its hardware to an applied load of 200 pounds. 5.1.2 Test Procedure The shelf was removed and weight was placed on top of the cabinet to prevent it from overturning. A cabinet door was opened 90˚ and a sling with four 50 pound weights was hung over the top of the door, at a point out from the hinge centerline. The door was then slowly moved through the full cycle of the hinge (up to a 160˚ arc). The weight was then removed, the door was swung through its full intended range of motion, and closed. 5.1.3 Acceptance Level There was no significant permanent distortion and the door operated normally after the weight was removed.
DOOR IMPACT TEST
5.2.1 Purpose of Test The test demonstrates the resistance to a 240 inch-pound impact to the door face. 5.2.2 Test Procedure Eight, 50 pound solid steel bars were placed on top of the cabinet to prevent overturn- ing. A 20 pound sand bag was then suspended from a pendulum support and dropped providing an impact of 240 inch-pound at the center of a closed door. 5.2.3 Acceptance Level Door and catch operated normally and show no signs of permanent damage.
DOOR CYCLE TEST
5.3.1 Purpose of Test The test demonstrates the ability of the door hinge hardware to withstand 100,000 cycles as a reliable measure of longevity. 5.3.2 Test Procedure A cycling mechanism, that swings a door through an arc of 90˚, was operated for 100,000 cycles. 5.3.3 Acceptance Level The door operated for the full 100,000 cycles without deterioration that significantly affected the function of the door. After completion of the test, the door operates freely and without binding.
DRAWER STATIC LOAD TEST
6.1.1 Purpose of Test The test demonstrates the ability to support a point load given to the front of the drawer and will challenge the attachment of the drawer head to the drawer. 6.1.2 Test Procedure Eight, 50 pound solid steel bars were placed on to of the cabinet to prevent overturning. A drawer was opened to 13” of travel and 150 pounds was hung from the drawer head, at the centerline of the drawer, for 5 minutes. The weight was then removed and the drawer was operated through the full cycle. 6.1.3 Acceptance Level There was no interference with the normal operation of the drawer.
DiversifiedCasework.com | info@DiversifiedCasework.com | 877.348.9663
SEFA Results
DRAWER & DOOR PULL TEST
6.2.1 Purpose of Test The test evaluates the strength of the pull and the pull hardware. 6.2.2 Test Procedure
The pulls were installed in accordance with Diversified Woodcrafts practice using spec- ified attaching hardware and method. A drawer and a door were blocked closed and a cable, pulley, and weight assembly was used to apply a force of 50 pounds perpendicular to each pull. Similar set-up was then used to hang a 50 pound weight from each pull. 6.2.3 Acceptance Level Pulls resisted the force and supported the weight and there is no evidence of permanent distortion.
DRAWER IMPACT TEST
6.3.1 Purpose of Test The test demonstrates the resistance to impact of the drawer bottom and the slide mecha- nism. 6.3.2 Test Procedure A drawer was opened to 13" of travel and then a 10 pound sand bag was dropped, from a height of 24", into the bottom of the drawer, at the centerline of the width of the drawer and 6" back from the inside face of the drawer. 6.3.3 Acceptance Level The drawer was operated through a full cycle and it operated normally. There was no visible deformation of the drawer. 6.4.1 Purpose of Test The test evaluates the strength of the drawer head, bottom, and back as a result of opening and closing the drawer with a rolling load. 6.4.2 Test Procedure The drawer was positioned on a table at a 45˚ angle and a 2" diameter by 12" long steel rod, that weighs approximately 10 pounds, was positioned 13" from the target impact area such that the rod rolled freely to impact the back of the drawer. The back was subjected to three impacts and then the drawer was reversed and the front was subject to three additional impacts. 6.4.3 Acceptance Level There are no visible signs of permanent damage. All joinery is intact and the drawer operat- ed normally when returned to the unit. DRAWER INTERNAL ROLLING IMPACT 6.5.1 Purpose of Test The test is intended to replicate years of operation of a drawer under full load. 6.5.2 Test Procedure A static load of 100 pounds (using ten 10 pound sand bags) was uniformly distributed in a drawer and the force required to activate the drawer was measured. The drawer was then opened from a closed position to a position within 1/4" of full extension and then closed. The procedure was repeated 50,000 times at a rate that did not exceed 10 cycles per minute. 6.5.3 Acceptance Level After the 50,000 cycles were completed, the drawer operated freely with no evidence of dragging, rubbing, or binding. The force required to open and close the loaded drawer increased by 7.1% which is less than the allowable maximum (20%). The force required to open and close the loaded drawer was less than 8 pounds. DRAWER CYCLE TEST
DiversifiedCasework.com | info@DiversifiedCasework.com | 877.348.9663
SEFA Results
SHELVING
7.1.1 Purpose of Test The test demonstrates the ability of a shelf and mounting hardware to support normal laboratory loads. 7.1.2 Test Procedure The shelf was mounted in the manner in which it was designed and a dial indicator was positioned under the center of the shelf and adjusted to zero. Thirteen 10 pound sand bags were uniformly distributed on the shelf to obtain a loading of 40 pounds per square foot of shelf area and the deflection was recorded. The weight was then removed from the shelf. 7.1.3 Acceptance Level The deflection of the shelf under load - 0.245” which is less than the allowable maximum of 0.25.” There was no significant permanent distortion of the shelf. 8.1.1 Purpose of Test The test demonstrates the strength of the back of the wall cabinet as well as the joinery of the cabinet and function of the doors when the unit is subjected to loads normally expected for laboratory furniture. 8.1.2 Test Procedure 130 pounds of sand bags, 40 pounds per square foot, were used to load the cabinet bottom, the cabinet top and shelves. With the weight in place the doors were operated through their full travel to verify normal operation. The weight was then removed and the doors were again operated through their full travelto verify normal operation. 8.1.3 Acceptance Level Doors opened and closed normally with and without the weight in place. There is no significant permanent deflection or damage to cabinet, cabinet back, cabinet top, cabinet bottom or shelf. WALL CABINET LOADING TEST 9.1.1 Purpose of Test This test will demonstrate the structural integrity of the table construction when subjected to a racking load. Most racking failures occur upon dragging an unloaded table across a floor. The ability of a table to resist a racking load will indicate less damage to the structure. The following tests were based on and adapted from ANSI/BIFMA X5.5-1989 American National Standard for Office Furnishings “Desk Products-Tests.” Adjustments have been made to better accommodate the specific applications of tables used in laboratories. 9.1.2 Test Procedure The table shall have a common two-by-four wood rail clamped on the centerline of the top parallel to the ends of the table. The table shall then be positioned at 45º, with one pair of legs on the floor and the other raised and supported (see Figure 14). The table shall then have 250 pounds (113.398 Kg) of weight (five 50-pound (22.679 Kg) bars) placed on its top and held in place by the two-by-four wood rail. The unit shall remain in this position for seventy-two hours. The unit shall be lowered without shock to the leveled surface and the general operation of the drawers shall be evaluated. 9.1.3 Acceptance Level When returned to normal position, the operation of the table was normal, and there was no permanent damage. TABLE TEST
DiversifiedCasework.com | info@DiversifiedCasework.com | 877.348.9663
SEFA Results
CHEMICAL RESITANCE TESTING ( RATING SCALE )
LEVEL 0 = NO DETECTABLE CHANGE LEVEL 1 = SLIGHT CHANGE IN COLOR OR GLOSS LEVEL 2 = SLIGHT SURFACE ETCHING OR SEVERE STAINING LEVEL 3 = PITTING, CRATERING, SWELLING, EROSION OF COATING, OBVIOUS & SIGNIFICANT SRETIORATION
CHEMICAL
RATING
COMMENTS
Amyl Acetate Ethyl Acetate
0 0 0 0 0 0 0 0 2 0 0 0 0 0 1 0 0 0 0 0 0 0 0 2 0 2 0 0 0 0 0 0 2 0 0 0 2 2 2 0 0 0 2 2 1 0 0 0 0
Acetic Acid 98%
Acetone
Acid Dichromate 5%
Butyl Alcohol Ethyl Alcohol Methyl Alcohol
Ammonium Hydroxide 28%
Severe staining (dark became black)
Benzene
Carbon Tetrachloride
Chloroform
Chromic Acid 60%
Cresol
Dichloro Acetic Acid Dimethylformanide
Slight change in gloss
Dioxane
Ethyl Ether
Formaldehyde 37% Formic Acid 90%
Furfural Gasoline
Hydrochloric Acid 37% Hydrofluoric Acit 48% Hydrogen Peroxide 3% Tincture of Iodine Methyl Ethyl Ketone Methylene Chloride Mono Chlorobenzene
Severe staining (surface became white)
Severe staining (surface became yellow)
VM&P
Nitric Acid 20% Nitric Acid 30% Nitric Acid 70%
Severe staining (surface became brown)
Phenol 90%
Phosphoric Acid 85% Silver Nitrate (Saturated) Sodium Hydroxide 10% Sodium Hydroxide 20% Sodium Hydroxide 40% Sodium Hydroxide Flake Sodium Sulfide (Saturated)
Severe staining (dark grain became black) Severe staining (dark grain became black) Severe staining (dark grain became black)
Sulfuric Acid 25% Sulfuric Acid 85% Sulfuric Acid 96%
Severe staining (dark grain became brown) Severe staining (dark grain became black) Slight change in color (yellow stain)
Sulfuric Acid 85% & Nitric Acid 70% (Equal Parts)
Toluene
Trichloroethylene
Xylene
Zinc Chloride Saturated TOTAL OF ALL RATINGS
20
DiversifiedCasework.com | info@DiversifiedCasework.com | 877.348.9663
SEFA Results
FINISH TEST
HOT WATER TEST
10.1.1 Purpose of Test The purpose of the chemical spot test is to evaluate the resis- tance the finish has to chemical spills. 10.1.2 Test Procedure The received sample measuring 14” x 24” was tested for chemical resistance as described herein. The panel was placed on a flat surface, cleaned with soap and water, then blotted dry. The panel was conditioned for 48 hours at 73+3˚F and 50+ 5% relative humidity. The panel was then tested for chemical resistance using forty-nine different chemical reagents by one of the following methods. Method A - Volatile chemicals were tested by placing a cotton ball saturated with reagent in the mouth of a 1-oz. bottle an inverting the bottle on the surface of the panel. Method B - Non-volatile chemicals were tested by placing five drops of the reagent on the surface of the panel and covering with a 24mm watch glass, convex side down. For both meth- ods, the reagents were left on the panel for a period of one hour, then washed off with water, cleaned with detergent and naptha, and rinsed with deionized water. The panel was then dried with a towel and evaluated after 24-hours at 73± 3˚F and 50 ± 5% relative humidity. 10.1.3 Acceptance Level Chemical test exceeded the requirements of no more than four of 49 chemicals tested failed with a fail rate of three.
10.2.1 Purpose of Test The purpose of this test is to insure the coating is resistant to hot water. 10.2.2 Test Procedure Hot water (190˚F. to 205˚F) was allowed to trickle with a steady stream and at a rate of not less than 6 ounces/minute on the finished surface, which was set at an angle of 45˚, for a period of five minutes. 10.3.3 Acceptance Level After cooling and wiping dry, the finish showed no visible effect from the hot water. 10.3.1 Purpose of Test The purpose of this test is to evaluate the ductility of the coating. 10.3.2 Test Procedure A one-pound ball approximately 2” in diameter was dropped from a distance of 12” onto the flat horizontal surface, which was coated to DWI standard manufacturing method. 10.3.3 Acceptance Level There is no visual evidence to the naked eye of cracks or checks in the finish due to impact. IMPACT TEST