FC1000E/SF1000E/X1020/L1020: EPA 2020 Certification Report

ATTACHMENT TO CALIBRATION CERTIFICATE 84106 AS LEFT DATA Page 3 of 3

Reading

Lower Limit

Measured Upper Limit

Measurement Uncertainty (k=2)

Error,

CMC

From of Meter

Reading From of Meter

STATUS

Standard,

Reading,

Meter,

Reading,

(k=2)

slpm 1.956 3.975 5.968 7.969 9.953

slpm 1.856 3.875 5.868 7.869 9.853

slpm 2.002 4.004 6.002 8.001 9.999

slpm slpm slpm slpm

2.056 4.075 6.068 8.069

0.046 0.029 0.034 0.032 0.046

0.010 0.020 0.030 0.040 0.050

0.01 0.02 0.03 0.04 0.05

Pass Pass Pass Pass Pass

10.053

ERROR CHART ( Inlet Pressure = 10 psig )

0 10 20 30 40 50 60 70 80 90 100

0.0

2.0

4.0

6.0

8.0

10.0

12.0

FLOW RATE ( Standard Reading )

slpm

Instrument Specifications

Meter's Calibrated Fluid:

Air Air 14.7 70 0 10 0.001 1.0

Test Fluid:

Meter's Standard Pressure: Meter's Standard Temperature:

slpm slpm ⁰F psia

Lower Range: Upper Range Resolution: Rated Accuracy:

% of Full Scale

Laboratory Ambient Conditions

Pressure: Humidity: Temperature:

psia %RH ⁰ F

14.39 52.17 70.02

Made with FlippingBook - professional solution for displaying marketing and sales documents online