ATTACHMENT TO CALIBRATION CERTIFICATE 84107 AS FOUND / AS LEFT DATA Page 2 of 2
Reading
Lower Limit
Measured Upper Limit
Measurement Uncertainty (k=2)
Error,
CMC
From of Meter
Reading From of Meter
STATUS
Standard,
Reading,
Meter,
Reading,
(k=2)
slpm 2.049 4.048 6.057 8.049
slpm 1.949 3.948 5.957 7.949 9.937
slpm 2.006 4.002 6.003 8.001 9.998
slpm slpm slpm slpm
2.149 4.148 6.157 8.149
-0.043 -0.046 -0.054 -0.048 -0.039
0.010 0.020 0.030 0.040 0.050
0.01 0.02 0.03 0.04 0.05
Pass Pass Pass Pass Pass
10.037
10.137
ERROR CHART ( Inlet Pressure = 10 psig )
-100 -90 -80 -70 -60 -50 -40 -30 -20 -10 0
0.0
2.0
4.0
6.0
8.0
10.0
12.0
FLOW RATE ( Standard Reading )
slpm
Instrument Specifications
Meter's Calibrated Fluid: Test Fluid:
Air Air 14.7 70 0 10 0.001 1.0 14.39 51.79 70.02
Meter's Standard Pressure: Meter's Standard Temperature:
slpm slpm ⁰F psia
Rated Accuracy: Lower Range: Upper Range Resolution:
% of Full Scale
psia %RH ⁰ F Laboratory Ambient Conditions
Pressure: Humidity: Temperature: