FC1000E/SF1000E/X1020/L1020: EPA 2020 Certification Report

Fox Valley Metrology 3114 Medalist Drive Oshkosh, WI 54902 (920) 426-5894 ● Fax (920) 426-8120 http://www.FoxValleyMetrology.com

CERTIFICATE OF CALIBRATION

Certificate No. ACT-1272

This certificate shall not be altered in any form or reproduced, except in full, without prior written approval from originating lab. These results relate only to the item(s) calibrated. Form Revision 6: 02/04/2012

Certification Number CK098-45054-574

Procedures Followed

FVE-020 rev. 2

Intertek 8431 Murphy Drive Middleton, WI 53562

For

Purchase Order #

Analog Scale Scales

Test Instrument

Standards Used

Instrument FVS-019D FVS-019G

Serial Number

Trace Number

Next Cal

26623

CI164-75132-439 CI247-38166-466 CJ247-29697-397

06/28/2020 09/28/2020 09/28/2020

Ohaus

Model Make

N/A N/A

Explorer E12140

FVS-710

B258010639

Identification Serial Number

█ ▀ ▀ ▀ █▄█▄ ██▄█ ▄▄ ▄█▄▄ █▄▀▀▀█▀ █ ▀▄ ██▄▄█▄ ▄█▄▄ █▄█ ▀▄▄█ █▄ █▄▄▄▄▄▄▄▄▄▄▄

713

Customer Location

Total expanded measurement uncertainties expressed are based on a confidence level of 95%; coverage factor of (k=2). The statement of compliance in this certificate was issued without taking the uncertainty of measurement into consideration. The customer shall assess the results and uncertainty when determining if the results meet their needs. (This is considered “shared responsibility.”) Uncertainties expressed in nominal units. All instruments have been calibrated against standards traceable to NIST. Calibration was completed in accordance with ISO/IEC 17025:2017, ANSI/NCSL Z540-1-1994 and ANSI/NCSL Z540.3-2006. Other standards listed upon request.

In Tolerance In Tolerance

Calibration Date Calibration Conditions Calibration Location Form Reviewed By Calibrated By Condition Returned Condition Received

William Trochinski M.T.

72.4°F, 22.4°C, 34.0%RH On Site

04/07/2020 10/07/2020

Recalibration Due

* denotes "Out of Tolerance"

Calibration Results

Feature

Nominal

Lower Limit

Upper Limit

As Found

As Left

Uncertainty

Increasing Load

(g)

(g)

(g)

(g)

(g)

(g)

5.0000 10.0000 20.0000 50.0000 100.0000 200.0000 210.0000

4.9999 9.9998 19.9998 49.9997 99.9997 199.9997 209.9997

5.0001 10.0002 20.0002 50.0003 100.0003 200.0003 210.0003

5.0000 10.0000 20.0000 50.0000 99.9999 199.9998 209.9998

5.0000 10.0000 20.0000 50.0000 99.9999 199.9998 209.9998

0.00007 0.00007 0.00006 0.00011 0.00023 0.00044 0.00046

Decreasing Load

(g)

(g)

(g)

(g)

(g)

(g)

20.0000 5.0000

19.9998 4.9999

20.0002 5.0001

19.9999 4.9999

19.9999 4.9999

0.00006 0.00006

Shift Test

(g)

(g)

(g)

(g)

(g)

(g)

Front

70.0000 70.0000 70.0000 70.0000

69.9997 69.9997 69.9997 69.9997

70.0003 70.0003 70.0003 70.0003

69.9999 69.9999 69.9998 69.9998

69.9999 69.9999 69.9998 69.9998

0.00016 0.00016 0.00016 0.00016

Left

Right Back

Comments Scale Capacity = 210 g; Precision = 0.0000 g; d = 0.0001 g; Class = I; Total Divisions = 2100000

Certification Number CK098-45054-574

Job Number JK070-36547-497

Page 1 of 1

Made with FlippingBook - professional solution for displaying marketing and sales documents online