4300 N Bell School Rd. Loves Park, IL 61111
www.creamcityscale.com
490 Enterprise Dr. Lake Mills, W I 53551
Certificate of Calibration
Acc red itation # 82374
Customer: Address: City, State Zip:
Certificate ID: ISO Number: Date
Intertek Testing Services 8431 Murphy Drive Middleton, WISCONSIN 53562
117041494600044201304 ISO/IEC 17025 82374 4/13/2020
Indicator Mfg.
Base Mfg. Rice Lake Base Model 4x4 HP-1K Base Serial
Cal Date 4/13/2020 Due Date 10/13/2020 Procedure QWI 6.4.2
Scale ID
Scale Location
Rice Lake
1134
Hearth
Indicator Model
Scale Class
Scale Range
520
III
0 - 1000 lb x 0.1 lb
Indicator Serial
Scale Status
Test Interval 1494600044
C42769
In Service
6 Months
EQUIPMENT CONDITIONS [Working] Non-Working Clean Dirty Out Of Level
SHIFT TEST Shift Test Result: [Pass] Fail Adjust Not Applicable
Position
Position
1
2
1
4
2
4
3
3
All tolerances calculated in conformance with Handbook 44 Table 6. LOAD TEST Preliminary Load Test Test Wt. Reading Error 0.0 lb
Final Load Test Test Wt.
Reading 0.0 lb 100.0 lb 250.0 lb 500.0 lb 1000.0 lb
Error 0.0 lb 0.0 lb 0.0 lb 0.0 lb 0.0 lb
0.0 lb 100.0 lb 250.0 lb 500.0 lb 1000.0 lb
0.0 lb 0.0 lb 0.0 lb 0.0 lb 0.0 lb
0.0 lb 100.0 lb 250.0 lb 500.0 lb 1000.0 lb
100.0 lb 250.0 lb 500.0 lb 1000.0 lb
In maintenance tolerance? [Yes] No N/A
In maintenance tolerance? [Yes] No N/A
TEST INFORMATION Test Weight Classification:
F
Traceability Certificate Number(s):
W18-217A Cal Date: 7/23/2018 Recal Date: 7/23/2020, W18-217B Cal Date: 7/23/2018 Recal Date: 7/23/2020
Standards Used:
1000 lb #169; 50 lb #71, 72, 73, 74, 75, 76, 77, 78, 79, 80
Expanded Uncertainty:
See Comments [Onsite] Offsite
Test Location: Overall Result:
[Pass] Fail Adjust Was the scale within customers required accuracy? [Yes] No N/A Environmental Conditions:
[Acceptable] Unacceptable
Temperature:
66°F 40%
Humidity:
Comments / Notes:
Measurement of Uncertainty: 100 lb.= .012 lb./ 250 lb.= .0295 lb./ 500 lb.= .059 lb./ 1000 lb.= .118 lb.
Technician:
Mark Baker
Scales were calibrated with certified test weights. Adjustments made to restore and/or maintain the accuracy of the scale conform to the tolerances established by the National Institute of Standards and Technology as specified in Handbook 44 Section 2.20, manufacturers specifications or other written agreement with customer. Best measurement of uncertainty calculated using a coverage factor of K=2. This provides confidence level of 95%. Acceptance rule w=0. Overall result of PASS or ADJUST indicates measurement below or equal to the acceptance limit: AL=TL. Overall result of FAIL indicates measurement result above acceptance limit AL=TL. This certificate shall not be reproduced, except in full, without the written approval of the laboratory. Measurement uncertainty available upon request.This calibration test is accredited and meets the requirements of ISO/IEC 17025:2017 & ANSI/NCSL Z540-1-1994 as verified by Perry Johnson Laboratory Accreditation. Refer to certificate and scope of accreditation 82374.
Page 1 of 1
Form: 7.8-02/10/26/2019