FC1000E/SF1000E/X1020/L1020: EPA 2020 Certification Report

Fox Valley Metrology 3114 Medalist Drive Oshkosh, WI 54902

CERTIFICATE OF CALIBRATION

(920) 426-5894 ● Fax (920) 426-8120 http://www.FoxValleyMetrology.com

Certificate No. ACT-1272

This certificate shall not be altered in any form or reproduced, except in full, without prior written approval from originating lab. These results relate only to the item(s) calibrated. Form Revision 6: 02/04/2012

Certification Number CK195-43111-502

Procedures Followed

FVE-060 rev. 0

Intertek 8431 Murphy Drive Middleton, WI 53562

For

Purchase Order #

Differential Pressure Transducer

Test Instrument

Standards Used

Instrument FVS-486 FVS-789 FVS-789A

Serial Number

Trace Number

Next Cal

N/A

CK104-51793-397 CK014-22597-466 CK014-22979-466

04/30/2021 02/28/2021 02/28/2021

Omega

Model Make

7543203 7461971

PX263-01D5V (Intertek)

X15160230

Identification Serial Number

1407

█ ▀ █ ▀ █▄█▄█▄ █▀▀ ▄▄█▀▄▄ ▄ █▄█▄█▄ ▀▄▄ ▄ █ ▀▀ ▄▀▄▄█▄▄ █▄▀ ▄ ▀▀█▄▄ █▀▀ ▀▄▄▀▀▄▀▄ █▄▄█▄▄▄██▄█▄▄▄

Customer Location

In Tolerance In Tolerance

Calibration Date Calibration Conditions Calibration Location Form Reviewed By Calibrated By Condition Returned Condition Received

Total expanded measurement uncertainties expressed are based on a confidence level of 95%; coverage factor of (k=2). The statement of compliance in this certificate was issued without taking the uncertainty of measurement into consideration. The customer shall assess the results and uncertainty when determining if the results meet their needs. (This is considered “shared responsibility.”) Uncertainties expressed in nominal units. All instruments have been calibrated against standards traceable to NIST. Calibration was completed in accordance with ISO/IEC 17025:2017, ANSI/NCSL Z540-1-1994 and ANSI/NCSL Z540.3-2006. Other standards listed upon request.

Christopher Moore M.T.

68.0°F, 20.0°C, 49.0%RH On Site

07/13/2020 01/13/2021

Recalibration Due

* denotes "Out of Tolerance"

Calibration Results

Feature

Nominal (inH20) -0.0000 -0.0500 -0.1000 -0.5000 -1.0000

Lower Limit

Upper Limit

As Found

As Left

Uncertainty

Pressure/Vacuum

(inH20) -0.0025 -0.0525 -0.1025 -0.5025 -1.0025 -0.0025 0.0475 0.0975 0.4975 0.9975

(inH20) 0.0025 -0.0475 -0.0975 -0.4975 -0.9975 0.0025 0.0525 0.1025 0.5025 1.0025

(inH20) 0.00007 0.00007 0.00007 0.00007 0.00006 0.00006 0.00006 0.00007 0.00007 0.00007

Low

0.0005 -0.0500 -0.1008 -0.5005 -1.0005 0.0009 0.0501 0.1005 0.5007 1.0001

0.0005 -0.0500 -0.1008 -0.5005 -1.0005 0.0009 0.0501 0.1005 0.5007 1.0001

Vacuum

High

0.0000 0.0500 0.1000 0.5000 1.0000

pressure

Certification Number CK195-43111-502

Job Number JK164-33231-497

Page 1 of 1