FC1000E/SF1000E/X1020/L1020: EPA 2020 Certification Report

ATTACHMENT TO CALIBRATION CERTIFICATE 82630 AS FOUND / AS LEFT DATA

Page 2 of 2

Reading

Lower Limit

Measured Upper Limit

Measurement Uncertainty (k=2)

CMC

From of Meter

Reading From of Meter

Error,

STATUS

Standard,

Reading,

Meter,

Reading,

(k=2)

Actual Air Velocity

fpm 270 522

fpm

fpm 250 524

fpm fpm

fpm

fpm STATUS

93

447 699

-20

6 9

5 9

Pass Pass Pass Pass Pass Pass Pass Pass

345

2

1574 2495 3570 4551 5520 5831

1397 2318 3393 4315 5284 5595

1591 2508 3567 4541 5518 5827

1751 2672 3747 4787 5756 6067

17 13

22 34 48 61 74 78

22 34 48 61 74 78

-3

-10

-2 -4

ERROR CHART

150

100

50

0

0

1000

2000

3000

4000

5000

6000

7000

-50

-100

-100

-150

fpm

AIR VELOCITY,

Instrument Specifications

Test Fluid: Lower Velocity Range: Upper Velocity Range: Velocity Resolution: Velocity Accuracy:

Air 80 5910 1

fpm fpm

+/-(3% Full Scale ≤3937fpm) +/-(4% Full Scale >3937fpm)

Laboratory Ambient Conditions

Pressure: Humidity: Temperature:

psia %RH ⁰ F

14.45 32.50 71.66